QRS subtraction for atrial electrograms: flat, linear and spline interpolations

The main objective of this article is to implement and compare QRS subtraction techniques for intracardiac atrial electrograms based on using the surface ECG as a reference. A band-pass filter between 8 and 20 Hz followed by rectification, and then a low-pass filter at 6 Hz are used for QRS detectio...

Full description

Saved in:
Bibliographic Details
Main Authors: Ahmad, Anita, Salinet Jr., J. L., Brown, P., Tuan, J. H., Stafford, P., Ng, G. Andre, Schlindwein, F. S.
Format: Article
Published: Springer-Verlag 2011
Subjects:
Online Access:http://eprints.utm.my/id/eprint/29385/
http://dx.doi.org/10.1007/s11517-011-0829-9
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The main objective of this article is to implement and compare QRS subtraction techniques for intracardiac atrial electrograms based on using the surface ECG as a reference. A band-pass filter between 8 and 20 Hz followed by rectification, and then a low-pass filter at 6 Hz are used for QRS detection. QRS subtraction was performed using three different approaches: flat, linear and spline interpolations. QRS subtraction affects the power of the signals but it normally does not affect the dominant frequency. The average power of the atrial electrograms after QRS subtraction is significantly reduced for frequencies above 10 Hz.