Optimisation Of Digestion Method For Determination Of Arsenic In Shrimp Paste Sample Using Atomic Absorption Spectrometry

The microwave digestion method was developed and verified for the determination of arsenic in shrimp paste samples. Experimental design for five factors (HNO3 and H2O2 volumes, sample weight, microwave power and digestion time) were used for the optimisation of sample digestion. For this purpose, tw...

全面介绍

Saved in:
书目详细资料
Main Authors: C. W. Zanariah, C. W. Ngah,, Mohd Adib, Yahya,
格式: Article
语言:English
出版: Elsevier Sci Ltd 2015
主题:
As
在线阅读:http://ddms.usim.edu.my/handle/123456789/8328
http://ac.els-cdn.com/S0308814612006693/1-s2.0-S0308814612006693-main.pdf?_tid=710a5286-aea7-11e5-bc6c-00000aacb361&acdnat=1451447168_26196fb85ada27bfbffd1732d82868f9
标签: 添加标签
没有标签, 成为第一个标记此记录!
实物特征
总结:The microwave digestion method was developed and verified for the determination of arsenic in shrimp paste samples. Experimental design for five factors (HNO3 and H2O2 volumes, sample weight, microwave power and digestion time) were used for the optimisation of sample digestion. For this purpose, two level half factorial design, which involves 16 experiments, was adopted. The concentration of arsenic was analysed by graphite furnace atomic absorption spectrometry. Design Expert(R) 7.0 software was used to interpret all data obtained. The combination of 2 mL HNO3 and 1 mL H2O2 volumes, 0.1 g sample weight, 1400W power and 5 min digestion time was found to be the optimum parameters required to digest the shrimp paste samples. Tests with spiked samples presented good recoveries with relative standard deviations between 0.32% and 5.35%. (C) 2012 Elsevier Ltd. All rights reserved.