Assessing the influence of neglected GC-FID variables on the multiple responses using multivariate optimization for the determination of ethanol and acetonitrile in radiopharmaceuticals

Analytical gas chromatography in line with a flame ionization detector (GC-FID) method was developed and validated for direct determination of organic solvents in [18F]fluoro-ethyl-tyrosine ([18F]FET), [18F]fluoromisonidazole ([18F]FMISO) and [18F]fluorothymidine ([18F]FLT). Variables of the splitle...

Full description

Saved in:
Bibliographic Details
Main Authors: Hassan, Hishar, Othman, Muhammad Faiz, Zakaria, Zainul Amiruddin, Ahmad Saad, Fathinul Fikri, Abdul Razak, Hairil Rashmizal
Format: Article
Published: Springer 2021
Online Access:http://psasir.upm.edu.my/id/eprint/96070/
https://link.springer.com/article/10.1007/s42452-021-04246-8
Tags: Add Tag
No Tags, Be the first to tag this record!
id my.upm.eprints.96070
record_format eprints
spelling my.upm.eprints.960702023-02-23T02:32:50Z http://psasir.upm.edu.my/id/eprint/96070/ Assessing the influence of neglected GC-FID variables on the multiple responses using multivariate optimization for the determination of ethanol and acetonitrile in radiopharmaceuticals Hassan, Hishar Othman, Muhammad Faiz Zakaria, Zainul Amiruddin Ahmad Saad, Fathinul Fikri Abdul Razak, Hairil Rashmizal Analytical gas chromatography in line with a flame ionization detector (GC-FID) method was developed and validated for direct determination of organic solvents in [18F]fluoro-ethyl-tyrosine ([18F]FET), [18F]fluoromisonidazole ([18F]FMISO) and [18F]fluorothymidine ([18F]FLT). Variables of the splitless time (min) and injection temperature (°C) on the response of analysis time and resolution were optimized with the assistance of a two-level full factorial design and desirability function of Derringer. The proposed procedure was validated following the International Council for Harmonisation of Technical Requirements for Pharmaceuticals for Human Use (ICH) Q2 (R1) guideline. Excellent linearity, R2 > 0.990, indicated that approximately 99% of the response variance could be predicted from ethanol and acetonitrile concentrations ranging from 0.5 to 6.0 mg mL−1 and 0.1 to 0.8 mg mL−1, respectively. The proposed procedure has proved to be selective, sensitive, and accurate (90–110%), with excellent repeatability and precision (RSD < 2%). In the robustness analysis, the findings from the calculated Standardized Effects Values (SE) were insignificant (p > 0.05) and demonstrated that the proposed method was robust for a splitless time of 1.0 ± 0.5 min and an injection temperature of 210 ± 10 °C. The proposed method was also successfully used for the quantitative determination of ethanol and acetonitrile in [18F]FET, [18F]FMISO, and [18F]FLT. Both solvents were well separated (R, 4.1–4.3) within 4.5 min. Therefore, the proposed method is relevant for routine quality control analysis of all 18F-radiopharmaceutical derivatives for the direct determination of ethanol and acetonitrile. Springer 2021 Article PeerReviewed Hassan, Hishar and Othman, Muhammad Faiz and Zakaria, Zainul Amiruddin and Ahmad Saad, Fathinul Fikri and Abdul Razak, Hairil Rashmizal (2021) Assessing the influence of neglected GC-FID variables on the multiple responses using multivariate optimization for the determination of ethanol and acetonitrile in radiopharmaceuticals. SN Applied Sciences, 3. art. no. 251. pp. 1-10. ISSN 2523-3971 https://link.springer.com/article/10.1007/s42452-021-04246-8 10.1007/s42452-021-04246-8
institution Universiti Putra Malaysia
building UPM Library
collection Institutional Repository
continent Asia
country Malaysia
content_provider Universiti Putra Malaysia
content_source UPM Institutional Repository
url_provider http://psasir.upm.edu.my/
description Analytical gas chromatography in line with a flame ionization detector (GC-FID) method was developed and validated for direct determination of organic solvents in [18F]fluoro-ethyl-tyrosine ([18F]FET), [18F]fluoromisonidazole ([18F]FMISO) and [18F]fluorothymidine ([18F]FLT). Variables of the splitless time (min) and injection temperature (°C) on the response of analysis time and resolution were optimized with the assistance of a two-level full factorial design and desirability function of Derringer. The proposed procedure was validated following the International Council for Harmonisation of Technical Requirements for Pharmaceuticals for Human Use (ICH) Q2 (R1) guideline. Excellent linearity, R2 > 0.990, indicated that approximately 99% of the response variance could be predicted from ethanol and acetonitrile concentrations ranging from 0.5 to 6.0 mg mL−1 and 0.1 to 0.8 mg mL−1, respectively. The proposed procedure has proved to be selective, sensitive, and accurate (90–110%), with excellent repeatability and precision (RSD < 2%). In the robustness analysis, the findings from the calculated Standardized Effects Values (SE) were insignificant (p > 0.05) and demonstrated that the proposed method was robust for a splitless time of 1.0 ± 0.5 min and an injection temperature of 210 ± 10 °C. The proposed method was also successfully used for the quantitative determination of ethanol and acetonitrile in [18F]FET, [18F]FMISO, and [18F]FLT. Both solvents were well separated (R, 4.1–4.3) within 4.5 min. Therefore, the proposed method is relevant for routine quality control analysis of all 18F-radiopharmaceutical derivatives for the direct determination of ethanol and acetonitrile.
format Article
author Hassan, Hishar
Othman, Muhammad Faiz
Zakaria, Zainul Amiruddin
Ahmad Saad, Fathinul Fikri
Abdul Razak, Hairil Rashmizal
spellingShingle Hassan, Hishar
Othman, Muhammad Faiz
Zakaria, Zainul Amiruddin
Ahmad Saad, Fathinul Fikri
Abdul Razak, Hairil Rashmizal
Assessing the influence of neglected GC-FID variables on the multiple responses using multivariate optimization for the determination of ethanol and acetonitrile in radiopharmaceuticals
author_facet Hassan, Hishar
Othman, Muhammad Faiz
Zakaria, Zainul Amiruddin
Ahmad Saad, Fathinul Fikri
Abdul Razak, Hairil Rashmizal
author_sort Hassan, Hishar
title Assessing the influence of neglected GC-FID variables on the multiple responses using multivariate optimization for the determination of ethanol and acetonitrile in radiopharmaceuticals
title_short Assessing the influence of neglected GC-FID variables on the multiple responses using multivariate optimization for the determination of ethanol and acetonitrile in radiopharmaceuticals
title_full Assessing the influence of neglected GC-FID variables on the multiple responses using multivariate optimization for the determination of ethanol and acetonitrile in radiopharmaceuticals
title_fullStr Assessing the influence of neglected GC-FID variables on the multiple responses using multivariate optimization for the determination of ethanol and acetonitrile in radiopharmaceuticals
title_full_unstemmed Assessing the influence of neglected GC-FID variables on the multiple responses using multivariate optimization for the determination of ethanol and acetonitrile in radiopharmaceuticals
title_sort assessing the influence of neglected gc-fid variables on the multiple responses using multivariate optimization for the determination of ethanol and acetonitrile in radiopharmaceuticals
publisher Springer
publishDate 2021
url http://psasir.upm.edu.my/id/eprint/96070/
https://link.springer.com/article/10.1007/s42452-021-04246-8
_version_ 1758952027491139584