Standardization of distance and angulation of light curing unit tip using distometer

The purpose of this study was to investigate the light intensity of selected light curing unit with varying distance and angulation of the light curing tip and lightmeter. Four types of light units; Spectrum 800 (Dentsply), Coltulux 3 (Coltene), Elipar FreeLight 2 (3M Espe) and Starlight Pro (Me...

Full description

Saved in:
Bibliographic Details
Main Authors: Radzi, Z., Abu Kasim, Noor Hayaty, Yahya, N.A., Abu Osman, Noor Azuan, Kassim, N.L.
Format: Conference or Workshop Item
Language:English
Published: 2007
Subjects:
Online Access:http://eprints.um.edu.my/2478/1/Standardization_of_distance_and_angulation_of_light_curing_unit_tip_using_distometer.pdf
http://eprints.um.edu.my/2478/
http://www.springerlink.com/content/g5457009364q656w/
Tags: Add Tag
No Tags, Be the first to tag this record!
id my.um.eprints.2478
record_format eprints
spelling my.um.eprints.24782019-02-07T08:20:22Z http://eprints.um.edu.my/2478/ Standardization of distance and angulation of light curing unit tip using distometer Radzi, Z. Abu Kasim, Noor Hayaty Yahya, N.A. Abu Osman, Noor Azuan Kassim, N.L. R Medicine R Medicine (General) RK Dentistry Practice of dentistry. Dental economics The purpose of this study was to investigate the light intensity of selected light curing unit with varying distance and angulation of the light curing tip and lightmeter. Four types of light units; Spectrum 800 (Dentsply), Coltulux 3 (Coltene), Elipar FreeLight 2 (3M Espe) and Starlight Pro (Mectron) were evaluated for light intensity at various distance between the light curing tip and the lightometer Cure Rite Denstply (0, 1, 3, 5, 10 and 15 mm). The light curing units were angulated at 45 degrees, 60 degrees and 90 degrees at a standardized 5 mm distance. The intensity of light curing is affected by the distance between the light curing tip and the lightmeter. However, the decrease in light intensity of the light curing unit was found not to obey the inverse square law for the distances 0 to 15 Mm. The study found that there was no significant difference between 45 degrees and 600 angulation between the light curing tip and the lightmeter. However, the decrease in light intensity is significant when compared to the light tip is placed perpendicular (90 degrees) to the aperture of the light meter. 2007 Conference or Workshop Item PeerReviewed application/pdf en http://eprints.um.edu.my/2478/1/Standardization_of_distance_and_angulation_of_light_curing_unit_tip_using_distometer.pdf Radzi, Z. and Abu Kasim, Noor Hayaty and Yahya, N.A. and Abu Osman, Noor Azuan and Kassim, N.L. (2007) Standardization of distance and angulation of light curing unit tip using distometer. In: 3rd Kuala Lumpur International Conference on Biomedical Engineering 2006, 2006, Kuala Lumpur. http://www.springerlink.com/content/g5457009364q656w/
institution Universiti Malaya
building UM Library
collection Institutional Repository
continent Asia
country Malaysia
content_provider Universiti Malaya
content_source UM Research Repository
url_provider http://eprints.um.edu.my/
language English
topic R Medicine
R Medicine (General)
RK Dentistry
Practice of dentistry. Dental economics
spellingShingle R Medicine
R Medicine (General)
RK Dentistry
Practice of dentistry. Dental economics
Radzi, Z.
Abu Kasim, Noor Hayaty
Yahya, N.A.
Abu Osman, Noor Azuan
Kassim, N.L.
Standardization of distance and angulation of light curing unit tip using distometer
description The purpose of this study was to investigate the light intensity of selected light curing unit with varying distance and angulation of the light curing tip and lightmeter. Four types of light units; Spectrum 800 (Dentsply), Coltulux 3 (Coltene), Elipar FreeLight 2 (3M Espe) and Starlight Pro (Mectron) were evaluated for light intensity at various distance between the light curing tip and the lightometer Cure Rite Denstply (0, 1, 3, 5, 10 and 15 mm). The light curing units were angulated at 45 degrees, 60 degrees and 90 degrees at a standardized 5 mm distance. The intensity of light curing is affected by the distance between the light curing tip and the lightmeter. However, the decrease in light intensity of the light curing unit was found not to obey the inverse square law for the distances 0 to 15 Mm. The study found that there was no significant difference between 45 degrees and 600 angulation between the light curing tip and the lightmeter. However, the decrease in light intensity is significant when compared to the light tip is placed perpendicular (90 degrees) to the aperture of the light meter.
format Conference or Workshop Item
author Radzi, Z.
Abu Kasim, Noor Hayaty
Yahya, N.A.
Abu Osman, Noor Azuan
Kassim, N.L.
author_facet Radzi, Z.
Abu Kasim, Noor Hayaty
Yahya, N.A.
Abu Osman, Noor Azuan
Kassim, N.L.
author_sort Radzi, Z.
title Standardization of distance and angulation of light curing unit tip using distometer
title_short Standardization of distance and angulation of light curing unit tip using distometer
title_full Standardization of distance and angulation of light curing unit tip using distometer
title_fullStr Standardization of distance and angulation of light curing unit tip using distometer
title_full_unstemmed Standardization of distance and angulation of light curing unit tip using distometer
title_sort standardization of distance and angulation of light curing unit tip using distometer
publishDate 2007
url http://eprints.um.edu.my/2478/1/Standardization_of_distance_and_angulation_of_light_curing_unit_tip_using_distometer.pdf
http://eprints.um.edu.my/2478/
http://www.springerlink.com/content/g5457009364q656w/
_version_ 1643686934218801152
score 13.160551