Standardization of distance and angulation of light curing unit tip using distometer

The purpose of this study was to investigate the light intensity of selected light curing unit with varying distance and angulation of the light curing tip and lightmeter. Four types of light units; Spectrum 800 (Dentsply), Coltulux 3 (Coltene), Elipar FreeLight 2 (3M Espe) and Starlight Pro (Me...

Full description

Saved in:
Bibliographic Details
Main Authors: Radzi, Z., Abu Kasim, Noor Hayaty, Yahya, N.A., Abu Osman, Noor Azuan, Kassim, N.L.
Format: Conference or Workshop Item
Language:English
Published: 2007
Subjects:
Online Access:http://eprints.um.edu.my/2478/1/Standardization_of_distance_and_angulation_of_light_curing_unit_tip_using_distometer.pdf
http://eprints.um.edu.my/2478/
http://www.springerlink.com/content/g5457009364q656w/
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:The purpose of this study was to investigate the light intensity of selected light curing unit with varying distance and angulation of the light curing tip and lightmeter. Four types of light units; Spectrum 800 (Dentsply), Coltulux 3 (Coltene), Elipar FreeLight 2 (3M Espe) and Starlight Pro (Mectron) were evaluated for light intensity at various distance between the light curing tip and the lightometer Cure Rite Denstply (0, 1, 3, 5, 10 and 15 mm). The light curing units were angulated at 45 degrees, 60 degrees and 90 degrees at a standardized 5 mm distance. The intensity of light curing is affected by the distance between the light curing tip and the lightmeter. However, the decrease in light intensity of the light curing unit was found not to obey the inverse square law for the distances 0 to 15 Mm. The study found that there was no significant difference between 45 degrees and 600 angulation between the light curing tip and the lightmeter. However, the decrease in light intensity is significant when compared to the light tip is placed perpendicular (90 degrees) to the aperture of the light meter.