An external fixator device is a medical implant used to keep fractured bones stabilized and in alignment. It consists of pins which are placed into the bone, extending outside the surface of the skin, and attached to a rigid external rod to keep it in place. The aim of this study is to investigate t...

Full description

Saved in:
Bibliographic Details
Main Authors: Mohammad Kata, Nurnedilah, Zainal Abidin, Nur Afikah, Abd. Aziz, Aishah Umairah, Abdullah, Abdul Halim, Ng, Bing Wui, Nasution, Ahmad Kafrawi, Abdul Kadir, Mohammed Rafiq, Ramlee, Muhammad Hanif
Format: Article
Language:English
Published: Penerbit UTM Press 2021
Subjects:
Online Access:http://eprints.utm.my/id/eprint/95384/1/NurnedilahMohammad2021_FiniteElementAnalysisofExternal.pdf
http://eprints.utm.my/id/eprint/95384/
http://dx.doi.org/10.11113/MJFAS.V17N3.2104
Tags: Add Tag
No Tags, Be the first to tag this record!
id my.utm.95384
record_format eprints
spelling my.utm.953842022-04-29T22:32:14Z http://eprints.utm.my/id/eprint/95384/ Finite element analysis of external fixator for treating femur fracture: analysis on stainless steel and titanium as material of external fixator Mohammad Kata, Nurnedilah Zainal Abidin, Nur Afikah Abd. Aziz, Aishah Umairah Abdullah, Abdul Halim Ng, Bing Wui Nasution, Ahmad Kafrawi Abdul Kadir, Mohammed Rafiq Ramlee, Muhammad Hanif Q Science (General) An external fixator device is a medical implant used to keep fractured bones stabilized and in alignment. It consists of pins which are placed into the bone, extending outside the surface of the skin, and attached to a rigid external rod to keep it in place. The aim of this study is to investigate the most suitable material used for the external fixator. Firstly, the 3D model of two unilateral uniplanar external fixator with the properties of titanium and stainless steel were constructed at Solidworks software with all the other parameters set to constant. Meanwhile, CT images of the lower limb were used to reconstruct a 3D model of the femur fracture at Mimics Medical software. Positioning and meshing of both the external fixator and the femur done at 3- Matics Medical and export as Patran for simulation at Marc Mentat software. 375 N load was applied at the most proximal femur to simulate stance phase of a gait cycle. From the findings, external fixator by using stainless steel as material properties have lower maximum von Mises Stress (18.40 MPa) at the femur and (103.69 MPa) at the fixator compared to the titanium (32.38 MPa) at the femur and (182.93 MPa) at the fixator. The result shows a difference of 75% of maximum von Mises Stress at the femur and the external fixator. Configuration by using stainless steel displaced 1.15 mm at the femur and 1.01 mm at the fixator which almost double value of displacement for titanium material for both femur (2.35 mm) and external fixator (2.11 mm). In conclusion, stainless steel external fixators provide better stability when compared to titanium external fixators. Penerbit UTM Press 2021-06 Article PeerReviewed application/pdf en http://eprints.utm.my/id/eprint/95384/1/NurnedilahMohammad2021_FiniteElementAnalysisofExternal.pdf Mohammad Kata, Nurnedilah and Zainal Abidin, Nur Afikah and Abd. Aziz, Aishah Umairah and Abdullah, Abdul Halim and Ng, Bing Wui and Nasution, Ahmad Kafrawi and Abdul Kadir, Mohammed Rafiq and Ramlee, Muhammad Hanif (2021) Finite element analysis of external fixator for treating femur fracture: analysis on stainless steel and titanium as material of external fixator. Malaysian Journal of Fundamental and Applied Sciences, 17 (3). pp. 274-284. ISSN 2289-599X http://dx.doi.org/10.11113/MJFAS.V17N3.2104 DOI:10.11113/MJFAS.V17N3.2104
institution Universiti Teknologi Malaysia
building UTM Library
collection Institutional Repository
continent Asia
country Malaysia
content_provider Universiti Teknologi Malaysia
content_source UTM Institutional Repository
url_provider http://eprints.utm.my/
language English
topic Q Science (General)
spellingShingle Q Science (General)
Mohammad Kata, Nurnedilah
Zainal Abidin, Nur Afikah
Abd. Aziz, Aishah Umairah
Abdullah, Abdul Halim
Ng, Bing Wui
Nasution, Ahmad Kafrawi
Abdul Kadir, Mohammed Rafiq
Ramlee, Muhammad Hanif
Finite element analysis of external fixator for treating femur fracture: analysis on stainless steel and titanium as material of external fixator
description An external fixator device is a medical implant used to keep fractured bones stabilized and in alignment. It consists of pins which are placed into the bone, extending outside the surface of the skin, and attached to a rigid external rod to keep it in place. The aim of this study is to investigate the most suitable material used for the external fixator. Firstly, the 3D model of two unilateral uniplanar external fixator with the properties of titanium and stainless steel were constructed at Solidworks software with all the other parameters set to constant. Meanwhile, CT images of the lower limb were used to reconstruct a 3D model of the femur fracture at Mimics Medical software. Positioning and meshing of both the external fixator and the femur done at 3- Matics Medical and export as Patran for simulation at Marc Mentat software. 375 N load was applied at the most proximal femur to simulate stance phase of a gait cycle. From the findings, external fixator by using stainless steel as material properties have lower maximum von Mises Stress (18.40 MPa) at the femur and (103.69 MPa) at the fixator compared to the titanium (32.38 MPa) at the femur and (182.93 MPa) at the fixator. The result shows a difference of 75% of maximum von Mises Stress at the femur and the external fixator. Configuration by using stainless steel displaced 1.15 mm at the femur and 1.01 mm at the fixator which almost double value of displacement for titanium material for both femur (2.35 mm) and external fixator (2.11 mm). In conclusion, stainless steel external fixators provide better stability when compared to titanium external fixators.
format Article
author Mohammad Kata, Nurnedilah
Zainal Abidin, Nur Afikah
Abd. Aziz, Aishah Umairah
Abdullah, Abdul Halim
Ng, Bing Wui
Nasution, Ahmad Kafrawi
Abdul Kadir, Mohammed Rafiq
Ramlee, Muhammad Hanif
author_facet Mohammad Kata, Nurnedilah
Zainal Abidin, Nur Afikah
Abd. Aziz, Aishah Umairah
Abdullah, Abdul Halim
Ng, Bing Wui
Nasution, Ahmad Kafrawi
Abdul Kadir, Mohammed Rafiq
Ramlee, Muhammad Hanif
author_sort Mohammad Kata, Nurnedilah
title Finite element analysis of external fixator for treating femur fracture: analysis on stainless steel and titanium as material of external fixator
title_short Finite element analysis of external fixator for treating femur fracture: analysis on stainless steel and titanium as material of external fixator
title_full Finite element analysis of external fixator for treating femur fracture: analysis on stainless steel and titanium as material of external fixator
title_fullStr Finite element analysis of external fixator for treating femur fracture: analysis on stainless steel and titanium as material of external fixator
title_full_unstemmed Finite element analysis of external fixator for treating femur fracture: analysis on stainless steel and titanium as material of external fixator
title_sort finite element analysis of external fixator for treating femur fracture: analysis on stainless steel and titanium as material of external fixator
publisher Penerbit UTM Press
publishDate 2021
url http://eprints.utm.my/id/eprint/95384/1/NurnedilahMohammad2021_FiniteElementAnalysisofExternal.pdf
http://eprints.utm.my/id/eprint/95384/
http://dx.doi.org/10.11113/MJFAS.V17N3.2104
_version_ 1732945466422722560
score 13.18916