Uncertainty of temperature measured by thermocouple

Reliability of data is important for researchers to verify their research result. For temperature measurement involving thermocouple, uncertainty needs to be determined before deciding the reliability of data. In this research, four error sources were proposed to have contributed in the uncertainty...

Full description

Saved in:
Bibliographic Details
Main Authors: Yoong, Yen Kee, Asako, Yutaka, Tan, Lit Ken, Che Sidik, Nor Azwadi
Format: Article
Language:English
Published: Penerbit Akademia Baru 2020
Subjects:
Online Access:http://eprints.utm.my/id/eprint/93640/1/YutakaAsako2020_UncertaintyofTemperaturemeasuredbyThermocouple.pdf
http://eprints.utm.my/id/eprint/93640/
http://dx.doi.org/10.37934/ARFMTS.68.1.5462
Tags: Add Tag
No Tags, Be the first to tag this record!
id my.utm.93640
record_format eprints
spelling my.utm.936402021-12-31T08:45:14Z http://eprints.utm.my/id/eprint/93640/ Uncertainty of temperature measured by thermocouple Yoong, Yen Kee Asako, Yutaka Tan, Lit Ken Che Sidik, Nor Azwadi T Technology (General) Reliability of data is important for researchers to verify their research result. For temperature measurement involving thermocouple, uncertainty needs to be determined before deciding the reliability of data. In this research, four error sources were proposed to have contributed in the uncertainty of temperature measured by a thermocouple which were resolution limit of data acquisition device, error in temperature measurement based on voltage measurement, reference junction compensation error and data fluctuation. Experiments were carried out to obtain reference junction compensation error and data fluctuation using HIOKI data logger (LR8400-20). The procedure to obtain the uncertainty of the measured temperature including the reference junction compensation uncertainty is proposed. The uncertainty was obtained by combining all the error values with root-sum-square equation. The uncertainty for K thermocouple obtained from this research was 0.42 °C. Penerbit Akademia Baru 2020 Article PeerReviewed application/pdf en http://eprints.utm.my/id/eprint/93640/1/YutakaAsako2020_UncertaintyofTemperaturemeasuredbyThermocouple.pdf Yoong, Yen Kee and Asako, Yutaka and Tan, Lit Ken and Che Sidik, Nor Azwadi (2020) Uncertainty of temperature measured by thermocouple. Journal of Advanced Research in Fluid Mechanics and Thermal Sciences, 68 (1). pp. 54-62. ISSN 2289-7879 http://dx.doi.org/10.37934/ARFMTS.68.1.5462
institution Universiti Teknologi Malaysia
building UTM Library
collection Institutional Repository
continent Asia
country Malaysia
content_provider Universiti Teknologi Malaysia
content_source UTM Institutional Repository
url_provider http://eprints.utm.my/
language English
topic T Technology (General)
spellingShingle T Technology (General)
Yoong, Yen Kee
Asako, Yutaka
Tan, Lit Ken
Che Sidik, Nor Azwadi
Uncertainty of temperature measured by thermocouple
description Reliability of data is important for researchers to verify their research result. For temperature measurement involving thermocouple, uncertainty needs to be determined before deciding the reliability of data. In this research, four error sources were proposed to have contributed in the uncertainty of temperature measured by a thermocouple which were resolution limit of data acquisition device, error in temperature measurement based on voltage measurement, reference junction compensation error and data fluctuation. Experiments were carried out to obtain reference junction compensation error and data fluctuation using HIOKI data logger (LR8400-20). The procedure to obtain the uncertainty of the measured temperature including the reference junction compensation uncertainty is proposed. The uncertainty was obtained by combining all the error values with root-sum-square equation. The uncertainty for K thermocouple obtained from this research was 0.42 °C.
format Article
author Yoong, Yen Kee
Asako, Yutaka
Tan, Lit Ken
Che Sidik, Nor Azwadi
author_facet Yoong, Yen Kee
Asako, Yutaka
Tan, Lit Ken
Che Sidik, Nor Azwadi
author_sort Yoong, Yen Kee
title Uncertainty of temperature measured by thermocouple
title_short Uncertainty of temperature measured by thermocouple
title_full Uncertainty of temperature measured by thermocouple
title_fullStr Uncertainty of temperature measured by thermocouple
title_full_unstemmed Uncertainty of temperature measured by thermocouple
title_sort uncertainty of temperature measured by thermocouple
publisher Penerbit Akademia Baru
publishDate 2020
url http://eprints.utm.my/id/eprint/93640/1/YutakaAsako2020_UncertaintyofTemperaturemeasuredbyThermocouple.pdf
http://eprints.utm.my/id/eprint/93640/
http://dx.doi.org/10.37934/ARFMTS.68.1.5462
_version_ 1720980103695958016
score 13.209306