Comparison between incident and reflected of two modes waves correlation with various porosities and thicknesses of bone phantom

The accuracy of the ultrasound measurement using through transmission technique to estimate bone quality based on fast and slow wave analysis is getting better but limited to the certain part of the skeletal site. Thus, the pulse echo technique with the application of fast and slow wave analysis is...

Full description

Saved in:
Bibliographic Details
Main Authors: Abd. Wahab, Muhamad Amin, Sudirman, Rubita, Abdul Razak, Mohd. Azhar, Mahmood, Nasrul Humaimi
Format: Article
Language:English
Published: Penerbit UTM Press 2020
Subjects:
Online Access:http://eprints.utm.my/id/eprint/91314/1/MuhamadAmin2020_ComparisonBetweenIncidentandReflectedofTwo.pdf
http://eprints.utm.my/id/eprint/91314/
https://mjfas.utm.my/index.php/mjfas/article/view/1930
Tags: Add Tag
No Tags, Be the first to tag this record!
id my.utm.91314
record_format eprints
spelling my.utm.913142021-06-30T12:07:37Z http://eprints.utm.my/id/eprint/91314/ Comparison between incident and reflected of two modes waves correlation with various porosities and thicknesses of bone phantom Abd. Wahab, Muhamad Amin Sudirman, Rubita Abdul Razak, Mohd. Azhar Mahmood, Nasrul Humaimi TK Electrical engineering. Electronics Nuclear engineering The accuracy of the ultrasound measurement using through transmission technique to estimate bone quality based on fast and slow wave analysis is getting better but limited to the certain part of the skeletal site. Thus, the pulse echo technique with the application of fast and slow wave analysis is proposed to solve the problem as the technique only uses one transducer to operate. The objective of this paper is to conduct experiments using through transmission and pulse echo technique to study the relationship of fast and slow wave with various porosity of three different thicknesses of polyurethane foam. Fast and slow waves will be separated from the "incident" and "reflected" wave (mix wave) by utilizing bandlimited deconvolution method. Then, the ultrasound parameters for mix, fast and slow waves for both incident and reflection waves are computed, plotted against porosity for each thickness and the results is compared in terms of the correlation coefficient. The overall results show that, the slope of attenuation of the incident fast waves shows a good relation, while the slope of attenuation of the reflected slow waves shows a good association with porosity for every thickness. Besides, the reflected wave may affected by scattering effect more compared to the incident wave, thus, cause the correlation between reflected wave parameters with porosity is lower. The overall results suggest, considering fast and slow wave to estimate bone quality might help to improve the measurement accuracy for both techniques. Penerbit UTM Press 2020-09 Article PeerReviewed application/pdf en http://eprints.utm.my/id/eprint/91314/1/MuhamadAmin2020_ComparisonBetweenIncidentandReflectedofTwo.pdf Abd. Wahab, Muhamad Amin and Sudirman, Rubita and Abdul Razak, Mohd. Azhar and Mahmood, Nasrul Humaimi (2020) Comparison between incident and reflected of two modes waves correlation with various porosities and thicknesses of bone phantom. Malaysian Journal of Fundamental and Applied Sciences, 16 (5). pp. 571-575. ISSN 2289-599X https://mjfas.utm.my/index.php/mjfas/article/view/1930
institution Universiti Teknologi Malaysia
building UTM Library
collection Institutional Repository
continent Asia
country Malaysia
content_provider Universiti Teknologi Malaysia
content_source UTM Institutional Repository
url_provider http://eprints.utm.my/
language English
topic TK Electrical engineering. Electronics Nuclear engineering
spellingShingle TK Electrical engineering. Electronics Nuclear engineering
Abd. Wahab, Muhamad Amin
Sudirman, Rubita
Abdul Razak, Mohd. Azhar
Mahmood, Nasrul Humaimi
Comparison between incident and reflected of two modes waves correlation with various porosities and thicknesses of bone phantom
description The accuracy of the ultrasound measurement using through transmission technique to estimate bone quality based on fast and slow wave analysis is getting better but limited to the certain part of the skeletal site. Thus, the pulse echo technique with the application of fast and slow wave analysis is proposed to solve the problem as the technique only uses one transducer to operate. The objective of this paper is to conduct experiments using through transmission and pulse echo technique to study the relationship of fast and slow wave with various porosity of three different thicknesses of polyurethane foam. Fast and slow waves will be separated from the "incident" and "reflected" wave (mix wave) by utilizing bandlimited deconvolution method. Then, the ultrasound parameters for mix, fast and slow waves for both incident and reflection waves are computed, plotted against porosity for each thickness and the results is compared in terms of the correlation coefficient. The overall results show that, the slope of attenuation of the incident fast waves shows a good relation, while the slope of attenuation of the reflected slow waves shows a good association with porosity for every thickness. Besides, the reflected wave may affected by scattering effect more compared to the incident wave, thus, cause the correlation between reflected wave parameters with porosity is lower. The overall results suggest, considering fast and slow wave to estimate bone quality might help to improve the measurement accuracy for both techniques.
format Article
author Abd. Wahab, Muhamad Amin
Sudirman, Rubita
Abdul Razak, Mohd. Azhar
Mahmood, Nasrul Humaimi
author_facet Abd. Wahab, Muhamad Amin
Sudirman, Rubita
Abdul Razak, Mohd. Azhar
Mahmood, Nasrul Humaimi
author_sort Abd. Wahab, Muhamad Amin
title Comparison between incident and reflected of two modes waves correlation with various porosities and thicknesses of bone phantom
title_short Comparison between incident and reflected of two modes waves correlation with various porosities and thicknesses of bone phantom
title_full Comparison between incident and reflected of two modes waves correlation with various porosities and thicknesses of bone phantom
title_fullStr Comparison between incident and reflected of two modes waves correlation with various porosities and thicknesses of bone phantom
title_full_unstemmed Comparison between incident and reflected of two modes waves correlation with various porosities and thicknesses of bone phantom
title_sort comparison between incident and reflected of two modes waves correlation with various porosities and thicknesses of bone phantom
publisher Penerbit UTM Press
publishDate 2020
url http://eprints.utm.my/id/eprint/91314/1/MuhamadAmin2020_ComparisonBetweenIncidentandReflectedofTwo.pdf
http://eprints.utm.my/id/eprint/91314/
https://mjfas.utm.my/index.php/mjfas/article/view/1930
_version_ 1705056696003461120
score 13.214268