Scatter radiation in the fluoroscopy-guided interventional room

Radiation scattered throughout the room during fluoroscopy-guided interventional (FGI) procedures was quantified at different locations using nanoDot optically stimulated luminescence dosemeters. All the tube angulation imaging shows that the radiation spectrum resembled a single peak distribution....

全面介绍

Saved in:
书目详细资料
Main Authors: Bohari, Asmah, Hashim, Suhairul, Mohd. Mustafa, Siti Norsyafiqah
格式: Article
出版: Oxford University Press 2020
主题:
在线阅读:http://eprints.utm.my/id/eprint/90800/
http://dx.doi.org/10.1093/rpd/ncz299
标签: 添加标签
没有标签, 成为第一个标记此记录!
实物特征
总结:Radiation scattered throughout the room during fluoroscopy-guided interventional (FGI) procedures was quantified at different locations using nanoDot optically stimulated luminescence dosemeters. All the tube angulation imaging shows that the radiation spectrum resembled a single peak distribution. The left anterior oblique 90° shows the highest single peak distribution (28.65 mSv/h). The single peak distribution for standard anteroposterior, left anterior oblique 45° and right anterior oblique 45° imaging was 13.32, 22.99 and 17.40 mSv/h, respectively. All tube angulation shows that the position of the interventional radiologist experienced a higher radiation level compared to other staffs. The doses of radiation varied widely around the perimeter of the patient's table and changed in accordance to imaging angles during procedures. Knowledge pertaining to radiation exposure levels is integral in order to avoid adverse risks, particularly among staff.