Surface analysis of retrieved bilateral UHMWPE tibial inserts under varus malalignment condition

Failure analysis on a retrieved ultra-high molecular weight (UHMWPE) knee tibial inserts of bilateral total knee replacement (TKR) was performed due to aseptic loosening detected after 16 years (left) and 12 years (right) in vivo services. Despite long implantation time, the effect of varus malalign...

Full description

Saved in:
Bibliographic Details
Main Authors: Khadijah, S. M. P., S. Liza, S. Liza, Shahemi, N., Merican, A. M., Abbas, A. A., Ayob, K. A.
Format: Article
Published: Elsevier Ltd 2020
Subjects:
Online Access:http://eprints.utm.my/id/eprint/90141/
http://dx.doi.org/10.1016/j.engfailanal.2020.104850
Tags: Add Tag
No Tags, Be the first to tag this record!
id my.utm.90141
record_format eprints
spelling my.utm.901412021-03-31T06:21:46Z http://eprints.utm.my/id/eprint/90141/ Surface analysis of retrieved bilateral UHMWPE tibial inserts under varus malalignment condition Khadijah, S. M. P. S. Liza, S. Liza Shahemi, N. Merican, A. M. Abbas, A. A. Ayob, K. A. T Technology (General) Failure analysis on a retrieved ultra-high molecular weight (UHMWPE) knee tibial inserts of bilateral total knee replacement (TKR) was performed due to aseptic loosening detected after 16 years (left) and 12 years (right) in vivo services. Despite long implantation time, the effect of varus malalignment present on a 71 years old female patient (body mass index, 35.1) with a non-active lifestyle will be considered as a factor towards the TKR failure. We, therefore, determined whether implant malalignment was associated with increased surface damages in both retrieved tibial inserts. Surface damage morphology was assessed using a 3D laser microscope and Scanning Electron Microscope (SEM). ATR-Fourier Transform Infra-Red (ATR-FTIR), Differential Scanning Calorimetry (DSC) and Gel-Permeation Chromatography (GPC) were used to measure changes of chemical and physical properties of retrieved inserts. Results show left-16 years insert possesses more severe wear degradation (crater and cracks) compare to wear on right-12 years insert (delamination, multidirectional scratches, and ripple). The surface roughness on the medial compartment seems to be higher than the lateral side for both inserts which can be affected by uneven load distribution contribute by the varus deformity. Higher crystallinity of left-16 years insert (66.99%) compare to right-12 years insert (56.52%) were an indicator of major mechanical changes happen on left insert which was contributed by oxidation with respect to implantation time of both inserts. Our findings revealed that in vivo oxidation is a main contributing factor to the failure of implants, but not varus malalignment. The material properties in the oxidized layer are significantly altered, including a very substantial reduction in molecular weight displayed by both inserts. Elsevier Ltd 2020-12 Article PeerReviewed Khadijah, S. M. P. and S. Liza, S. Liza and Shahemi, N. and Merican, A. M. and Abbas, A. A. and Ayob, K. A. (2020) Surface analysis of retrieved bilateral UHMWPE tibial inserts under varus malalignment condition. Engineering Failure Analysis, 118 . p. 104850. ISSN 1350-6307 http://dx.doi.org/10.1016/j.engfailanal.2020.104850
institution Universiti Teknologi Malaysia
building UTM Library
collection Institutional Repository
continent Asia
country Malaysia
content_provider Universiti Teknologi Malaysia
content_source UTM Institutional Repository
url_provider http://eprints.utm.my/
topic T Technology (General)
spellingShingle T Technology (General)
Khadijah, S. M. P.
S. Liza, S. Liza
Shahemi, N.
Merican, A. M.
Abbas, A. A.
Ayob, K. A.
Surface analysis of retrieved bilateral UHMWPE tibial inserts under varus malalignment condition
description Failure analysis on a retrieved ultra-high molecular weight (UHMWPE) knee tibial inserts of bilateral total knee replacement (TKR) was performed due to aseptic loosening detected after 16 years (left) and 12 years (right) in vivo services. Despite long implantation time, the effect of varus malalignment present on a 71 years old female patient (body mass index, 35.1) with a non-active lifestyle will be considered as a factor towards the TKR failure. We, therefore, determined whether implant malalignment was associated with increased surface damages in both retrieved tibial inserts. Surface damage morphology was assessed using a 3D laser microscope and Scanning Electron Microscope (SEM). ATR-Fourier Transform Infra-Red (ATR-FTIR), Differential Scanning Calorimetry (DSC) and Gel-Permeation Chromatography (GPC) were used to measure changes of chemical and physical properties of retrieved inserts. Results show left-16 years insert possesses more severe wear degradation (crater and cracks) compare to wear on right-12 years insert (delamination, multidirectional scratches, and ripple). The surface roughness on the medial compartment seems to be higher than the lateral side for both inserts which can be affected by uneven load distribution contribute by the varus deformity. Higher crystallinity of left-16 years insert (66.99%) compare to right-12 years insert (56.52%) were an indicator of major mechanical changes happen on left insert which was contributed by oxidation with respect to implantation time of both inserts. Our findings revealed that in vivo oxidation is a main contributing factor to the failure of implants, but not varus malalignment. The material properties in the oxidized layer are significantly altered, including a very substantial reduction in molecular weight displayed by both inserts.
format Article
author Khadijah, S. M. P.
S. Liza, S. Liza
Shahemi, N.
Merican, A. M.
Abbas, A. A.
Ayob, K. A.
author_facet Khadijah, S. M. P.
S. Liza, S. Liza
Shahemi, N.
Merican, A. M.
Abbas, A. A.
Ayob, K. A.
author_sort Khadijah, S. M. P.
title Surface analysis of retrieved bilateral UHMWPE tibial inserts under varus malalignment condition
title_short Surface analysis of retrieved bilateral UHMWPE tibial inserts under varus malalignment condition
title_full Surface analysis of retrieved bilateral UHMWPE tibial inserts under varus malalignment condition
title_fullStr Surface analysis of retrieved bilateral UHMWPE tibial inserts under varus malalignment condition
title_full_unstemmed Surface analysis of retrieved bilateral UHMWPE tibial inserts under varus malalignment condition
title_sort surface analysis of retrieved bilateral uhmwpe tibial inserts under varus malalignment condition
publisher Elsevier Ltd
publishDate 2020
url http://eprints.utm.my/id/eprint/90141/
http://dx.doi.org/10.1016/j.engfailanal.2020.104850
_version_ 1696976267378688000