Numerical study of a left ventricular assist device (LVAD) with different blade heights and tip clearances
One treatment modality for heart failure is to employ a mechanical heart assist device to increase blood flow to peripheral organs. There are various kinds of axial and centrifugal type mechanical pumps available for implantation depending on patient condition. Axial pumps are smaller in size compar...
Saved in:
Main Authors: | , , , , , , , |
---|---|
Format: | Article |
Published: |
Universiti Tun Hussein Onn Malaysia
2018
|
Subjects: | |
Online Access: | http://eprints.utm.my/id/eprint/84302/ https://publisher.uthm.edu.my |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
id |
my.utm.84302 |
---|---|
record_format |
eprints |
spelling |
my.utm.843022019-12-28T01:46:32Z http://eprints.utm.my/id/eprint/84302/ Numerical study of a left ventricular assist device (LVAD) with different blade heights and tip clearances Md. Khudzari, Ahmad Zahran Osman, Kahar Padzillah, Muhamad Hasbullah Harun, Muhamad Noor Abdul Kadir, Muhammad Rashidi Abdul Aziz, Mohammad Azim Abdul Rahim, Aizai Azan Abdul Taib, Mohamed Ezani TJ Mechanical engineering and machinery One treatment modality for heart failure is to employ a mechanical heart assist device to increase blood flow to peripheral organs. There are various kinds of axial and centrifugal type mechanical pumps available for implantation depending on patient condition. Axial pumps are smaller in size comparatively, although centrifugal pumps have the advantages of lower rotational speed as well as better maintaining any native blood flow pulsatility. This work presents the results of the numerical study of the centrifugal blood pump configured as a Left Ventricular Assist Device (LVAD). The pump design utilized standard industrial centrifugal pump design principles but applied to smaller sized blood pumps. Flow characteristics are modelled using 3-dimensional steady state models operating at design speed of 2000 rpm using Newtonian blood properties for the fluid. Two design parameters of the pump are studied, the impeller blade height and tip clearance resulting in nine model variants. Analysis includes the hydrodynamic performance of the pump and the flow characteristics in the pump. A haemolysis prediction model quantifying red blood cell stress from exposure time and shear stress was used for quantitative predictions of haemolysis within the blood pump. Blood damage estimation was calculated along each path-line and averaged to a single value. By using a ranked selection method, the model with the 15 mm blade height and 800 µm tip clearance was selected as the preferred configuration with Haemolysis Index of 0.01 mmHg, efficiency of 58% at 104 mmHg outlet pressure. Universiti Tun Hussein Onn Malaysia 2018 Article PeerReviewed Md. Khudzari, Ahmad Zahran and Osman, Kahar and Padzillah, Muhamad Hasbullah and Harun, Muhamad Noor and Abdul Kadir, Muhammad Rashidi and Abdul Aziz, Mohammad Azim and Abdul Rahim, Aizai Azan and Abdul Taib, Mohamed Ezani (2018) Numerical study of a left ventricular assist device (LVAD) with different blade heights and tip clearances. International Journal of Integrated Engineering, 10 (4). pp. 17-26. ISSN 2229-838X https://publisher.uthm.edu.my |
institution |
Universiti Teknologi Malaysia |
building |
UTM Library |
collection |
Institutional Repository |
continent |
Asia |
country |
Malaysia |
content_provider |
Universiti Teknologi Malaysia |
content_source |
UTM Institutional Repository |
url_provider |
http://eprints.utm.my/ |
topic |
TJ Mechanical engineering and machinery |
spellingShingle |
TJ Mechanical engineering and machinery Md. Khudzari, Ahmad Zahran Osman, Kahar Padzillah, Muhamad Hasbullah Harun, Muhamad Noor Abdul Kadir, Muhammad Rashidi Abdul Aziz, Mohammad Azim Abdul Rahim, Aizai Azan Abdul Taib, Mohamed Ezani Numerical study of a left ventricular assist device (LVAD) with different blade heights and tip clearances |
description |
One treatment modality for heart failure is to employ a mechanical heart assist device to increase blood flow to peripheral organs. There are various kinds of axial and centrifugal type mechanical pumps available for implantation depending on patient condition. Axial pumps are smaller in size comparatively, although centrifugal pumps have the advantages of lower rotational speed as well as better maintaining any native blood flow pulsatility. This work presents the results of the numerical study of the centrifugal blood pump configured as a Left Ventricular Assist Device (LVAD). The pump design utilized standard industrial centrifugal pump design principles but applied to smaller sized blood pumps. Flow characteristics are modelled using 3-dimensional steady state models operating at design speed of 2000 rpm using Newtonian blood properties for the fluid. Two design parameters of the pump are studied, the impeller blade height and tip clearance resulting in nine model variants. Analysis includes the hydrodynamic performance of the pump and the flow characteristics in the pump. A haemolysis prediction model quantifying red blood cell stress from exposure time and shear stress was used for quantitative predictions of haemolysis within the blood pump. Blood damage estimation was calculated along each path-line and averaged to a single value. By using a ranked selection method, the model with the 15 mm blade height and 800 µm tip clearance was selected as the preferred configuration with Haemolysis Index of 0.01 mmHg, efficiency of 58% at 104 mmHg outlet pressure. |
format |
Article |
author |
Md. Khudzari, Ahmad Zahran Osman, Kahar Padzillah, Muhamad Hasbullah Harun, Muhamad Noor Abdul Kadir, Muhammad Rashidi Abdul Aziz, Mohammad Azim Abdul Rahim, Aizai Azan Abdul Taib, Mohamed Ezani |
author_facet |
Md. Khudzari, Ahmad Zahran Osman, Kahar Padzillah, Muhamad Hasbullah Harun, Muhamad Noor Abdul Kadir, Muhammad Rashidi Abdul Aziz, Mohammad Azim Abdul Rahim, Aizai Azan Abdul Taib, Mohamed Ezani |
author_sort |
Md. Khudzari, Ahmad Zahran |
title |
Numerical study of a left ventricular assist device (LVAD) with different blade heights and tip clearances |
title_short |
Numerical study of a left ventricular assist device (LVAD) with different blade heights and tip clearances |
title_full |
Numerical study of a left ventricular assist device (LVAD) with different blade heights and tip clearances |
title_fullStr |
Numerical study of a left ventricular assist device (LVAD) with different blade heights and tip clearances |
title_full_unstemmed |
Numerical study of a left ventricular assist device (LVAD) with different blade heights and tip clearances |
title_sort |
numerical study of a left ventricular assist device (lvad) with different blade heights and tip clearances |
publisher |
Universiti Tun Hussein Onn Malaysia |
publishDate |
2018 |
url |
http://eprints.utm.my/id/eprint/84302/ https://publisher.uthm.edu.my |
_version_ |
1654960068716658688 |
score |
13.211869 |