Improvement in measurement accuracy for hybrid scanner

The capability to provide dense three-dimensional (3D) data (point clouds) at high speed and at high accuracy has made terrestrial laser scanners (TLS) widely used for many purposes especially for documentation, management and analysis. However, similar to other 3D sensors, proper understanding rega...

Full description

Saved in:
Bibliographic Details
Main Authors: Abbas, Mohamed Azwan, Setan, Halim, Majid, Zulkepli, Chong, Albert, Lichti, Derek D.
Format: Article
Language:English
Published: Institute of Physics Publishing 2014
Subjects:
Online Access:http://eprints.utm.my/id/eprint/53109/1/HalimSetan2014_Improvementinmeasurementaccuracy.pdf
http://eprints.utm.my/id/eprint/53109/
http://dx.doi.org/10.1088/1755-1315/18/1/012066
Tags: Add Tag
No Tags, Be the first to tag this record!
id my.utm.53109
record_format eprints
spelling my.utm.531092018-07-19T07:25:07Z http://eprints.utm.my/id/eprint/53109/ Improvement in measurement accuracy for hybrid scanner Abbas, Mohamed Azwan Setan, Halim Majid, Zulkepli Chong, Albert Lichti, Derek D. HD Industries. Land use. Labor The capability to provide dense three-dimensional (3D) data (point clouds) at high speed and at high accuracy has made terrestrial laser scanners (TLS) widely used for many purposes especially for documentation, management and analysis. However, similar to other 3D sensors, proper understanding regarding the error sources is necessary to ensure high quality data. A procedure known as calibration is employed to evaluate these errors. This process is crucial for TLS in order to make it suitable for accurate 3D applications (e.g. industrial measurement, reverse engineering and monitoring). Two calibration procedures available for TLS: 1) component, and 2) system calibration. The requirements of special laboratories and tools which are not affordable by most TLS users have become principle drawback for component calibration. In contrast, system calibration only requires a room with appropriate targets. By employing optimal network configuration, this study has performed system calibration through self-calibration for Leica ScanStation C10 scanner. A laboratory with dimensions of 15.5 m × 9 m × 3 m and 138 well-distributed planar targets were used to derive four calibration parameters. Statistical analysis (e.g. t-test) has shown that only two calculated parameters, the constant rangefinder offset error (0.7 mm) and the vertical circle index error (-45.4") were significant for the calibrated scanner. Photogrammetric technique was utilised to calibrate the 3D test points at the calibration field. By using the test points, the residual pattern of raw data and self-calibration results were plotted into the graph to visually demonstrate the improvement in accuracy for Leica ScanStation C10 scanner. Institute of Physics Publishing 2014 Article PeerReviewed application/pdf en http://eprints.utm.my/id/eprint/53109/1/HalimSetan2014_Improvementinmeasurementaccuracy.pdf Abbas, Mohamed Azwan and Setan, Halim and Majid, Zulkepli and Chong, Albert and Lichti, Derek D. (2014) Improvement in measurement accuracy for hybrid scanner. IOP Conference Series: Earth and Environmental Science, 18 (1). ISSN 1755-1315 http://dx.doi.org/10.1088/1755-1315/18/1/012066 DOI: 10.1088/1755-1315/18/1/012066
institution Universiti Teknologi Malaysia
building UTM Library
collection Institutional Repository
continent Asia
country Malaysia
content_provider Universiti Teknologi Malaysia
content_source UTM Institutional Repository
url_provider http://eprints.utm.my/
language English
topic HD Industries. Land use. Labor
spellingShingle HD Industries. Land use. Labor
Abbas, Mohamed Azwan
Setan, Halim
Majid, Zulkepli
Chong, Albert
Lichti, Derek D.
Improvement in measurement accuracy for hybrid scanner
description The capability to provide dense three-dimensional (3D) data (point clouds) at high speed and at high accuracy has made terrestrial laser scanners (TLS) widely used for many purposes especially for documentation, management and analysis. However, similar to other 3D sensors, proper understanding regarding the error sources is necessary to ensure high quality data. A procedure known as calibration is employed to evaluate these errors. This process is crucial for TLS in order to make it suitable for accurate 3D applications (e.g. industrial measurement, reverse engineering and monitoring). Two calibration procedures available for TLS: 1) component, and 2) system calibration. The requirements of special laboratories and tools which are not affordable by most TLS users have become principle drawback for component calibration. In contrast, system calibration only requires a room with appropriate targets. By employing optimal network configuration, this study has performed system calibration through self-calibration for Leica ScanStation C10 scanner. A laboratory with dimensions of 15.5 m × 9 m × 3 m and 138 well-distributed planar targets were used to derive four calibration parameters. Statistical analysis (e.g. t-test) has shown that only two calculated parameters, the constant rangefinder offset error (0.7 mm) and the vertical circle index error (-45.4") were significant for the calibrated scanner. Photogrammetric technique was utilised to calibrate the 3D test points at the calibration field. By using the test points, the residual pattern of raw data and self-calibration results were plotted into the graph to visually demonstrate the improvement in accuracy for Leica ScanStation C10 scanner.
format Article
author Abbas, Mohamed Azwan
Setan, Halim
Majid, Zulkepli
Chong, Albert
Lichti, Derek D.
author_facet Abbas, Mohamed Azwan
Setan, Halim
Majid, Zulkepli
Chong, Albert
Lichti, Derek D.
author_sort Abbas, Mohamed Azwan
title Improvement in measurement accuracy for hybrid scanner
title_short Improvement in measurement accuracy for hybrid scanner
title_full Improvement in measurement accuracy for hybrid scanner
title_fullStr Improvement in measurement accuracy for hybrid scanner
title_full_unstemmed Improvement in measurement accuracy for hybrid scanner
title_sort improvement in measurement accuracy for hybrid scanner
publisher Institute of Physics Publishing
publishDate 2014
url http://eprints.utm.my/id/eprint/53109/1/HalimSetan2014_Improvementinmeasurementaccuracy.pdf
http://eprints.utm.my/id/eprint/53109/
http://dx.doi.org/10.1088/1755-1315/18/1/012066
_version_ 1643653301212807168
score 13.19449