Design analysis of TMJ implant under physiological loading condition

The aim of this study was to determine the behavior of Temporomandibular joint implant made of different materials when it is subjected to loads resulted of daily movements. A three dimensional model of a lower mandible of a human were developed from a CT scan dataset. A commercial Temporomandibular...

Full description

Saved in:
Bibliographic Details
Main Authors: Arabshahi, Zohreh, Kashani, Jamal, Rahimian Koloor, Seyed Saeid, Abdul Kadir, Mohammed Rafiq, Azari, Abbas
Format: Article
Published: 2012
Subjects:
Online Access:http://eprints.utm.my/id/eprint/46779/
http://dx.doi.org/10.4028/www.scientific.net/AMR.488-489.996
Tags: Add Tag
No Tags, Be the first to tag this record!
id my.utm.46779
record_format eprints
spelling my.utm.467792017-09-19T08:49:35Z http://eprints.utm.my/id/eprint/46779/ Design analysis of TMJ implant under physiological loading condition Arabshahi, Zohreh Kashani, Jamal Rahimian Koloor, Seyed Saeid Abdul Kadir, Mohammed Rafiq Azari, Abbas TA Engineering (General). Civil engineering (General) The aim of this study was to determine the behavior of Temporomandibular joint implant made of different materials when it is subjected to loads resulted of daily movements. A three dimensional model of a lower mandible of a human were developed from a CT scan dataset. A commercial Temporomandibular joint implant and fixation screws were modeled. Three dimensional finite element models of implanted mandible analyzed under static loading of five clunching tasks. The investigation went to study on two materials of titanium and Cobalt-chromium alloy for implant. The equivalent strain of whole structure was computed and utilized for evaluation of implant design. The safety factor results showed that both materials were capable to carry the applied complex loads. 2012 Article PeerReviewed Arabshahi, Zohreh and Kashani, Jamal and Rahimian Koloor, Seyed Saeid and Abdul Kadir, Mohammed Rafiq and Azari, Abbas (2012) Design analysis of TMJ implant under physiological loading condition. Advanced Materials Research, 488-48 . pp. 996-1000. ISSN 1022-6680 http://dx.doi.org/10.4028/www.scientific.net/AMR.488-489.996
institution Universiti Teknologi Malaysia
building UTM Library
collection Institutional Repository
continent Asia
country Malaysia
content_provider Universiti Teknologi Malaysia
content_source UTM Institutional Repository
url_provider http://eprints.utm.my/
topic TA Engineering (General). Civil engineering (General)
spellingShingle TA Engineering (General). Civil engineering (General)
Arabshahi, Zohreh
Kashani, Jamal
Rahimian Koloor, Seyed Saeid
Abdul Kadir, Mohammed Rafiq
Azari, Abbas
Design analysis of TMJ implant under physiological loading condition
description The aim of this study was to determine the behavior of Temporomandibular joint implant made of different materials when it is subjected to loads resulted of daily movements. A three dimensional model of a lower mandible of a human were developed from a CT scan dataset. A commercial Temporomandibular joint implant and fixation screws were modeled. Three dimensional finite element models of implanted mandible analyzed under static loading of five clunching tasks. The investigation went to study on two materials of titanium and Cobalt-chromium alloy for implant. The equivalent strain of whole structure was computed and utilized for evaluation of implant design. The safety factor results showed that both materials were capable to carry the applied complex loads.
format Article
author Arabshahi, Zohreh
Kashani, Jamal
Rahimian Koloor, Seyed Saeid
Abdul Kadir, Mohammed Rafiq
Azari, Abbas
author_facet Arabshahi, Zohreh
Kashani, Jamal
Rahimian Koloor, Seyed Saeid
Abdul Kadir, Mohammed Rafiq
Azari, Abbas
author_sort Arabshahi, Zohreh
title Design analysis of TMJ implant under physiological loading condition
title_short Design analysis of TMJ implant under physiological loading condition
title_full Design analysis of TMJ implant under physiological loading condition
title_fullStr Design analysis of TMJ implant under physiological loading condition
title_full_unstemmed Design analysis of TMJ implant under physiological loading condition
title_sort design analysis of tmj implant under physiological loading condition
publishDate 2012
url http://eprints.utm.my/id/eprint/46779/
http://dx.doi.org/10.4028/www.scientific.net/AMR.488-489.996
_version_ 1643652137665691648
score 13.211869