Critical view on the usage of C-ring specimen for stress corrosion crack (SCC) test on orthopedic implant : experimental, numerical and analytical approaches
As the failures of orthopedic devices due to stress corrosion cracking (SCC) have become more frequent nowadays, research on this area also has become popular. Many published articles show the basic characterizations and evaluations of the SCC performed based on ASTM standards by using the C-ring sa...
Saved in:
Main Authors: | Prawoto, Yunan, Djuansjah, Joy Rizki Pangestu, Wan Nik, W. B., Enemuoh, Emmanuel |
---|---|
Format: | Article |
Published: |
2012
|
Subjects: | |
Online Access: | http://eprints.utm.my/id/eprint/46755/ https://dx.doi.org/10.1016/j.msec.2012.04.003 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Similar Items
-
Re-visiting the ‘rule of mixture’ used in materials with multiple constituting phases: A technical note on morphological considerations in austenite case study
by: Prawoto, Yunan, et al.
Published: (2012) -
Incorporating the morphological difference of corrosion cracks in computational fracture mechanics approach
by: Prawoto, Yunan
Published: (2012) -
Design and fabrication of an improved crack growth testing apparatus with stress corrosion cracking (SCC) study
by: Ibrahim Manai, Abdul Hakim
Published: (2006) -
Construction of a crack growth testing apparatus associated with stress corrosion cracking (SCC) study
by: Peter Anderson, anak Joseph
Published: (2005) -
Porous magnesium scaffolds for bone implant applications: a review
by: Noor Jasmawati, Noor Jasmawati, et al.
Published: (2015)