Alternative acoustic pulse echo immersion measurement system development for nonporous tissue mimicking materials

Acoustic properties are important to evaluate the compatibility of tested samples as tissue mimicking materials (TMMs). Common acoustic measurement systems require distilled water as the propagation medium. However, their accuracies are affected by the small change in the medium density and the inac...

Full description

Saved in:
Bibliographic Details
Main Author: Mat Daud, Anis Nazihah
Format: Thesis
Language:English
Published: 2020
Subjects:
Online Access:http://eprints.utm.my/id/eprint/101684/1/AnisNazihahMatDaudPFS2020.pdf
http://eprints.utm.my/id/eprint/101684/
http://dms.library.utm.my:8080/vital/access/manager/Repository/vital:145905
Tags: Add Tag
No Tags, Be the first to tag this record!
id my.utm.101684
record_format eprints
spelling my.utm.1016842023-07-09T00:46:22Z http://eprints.utm.my/id/eprint/101684/ Alternative acoustic pulse echo immersion measurement system development for nonporous tissue mimicking materials Mat Daud, Anis Nazihah QD Chemistry Acoustic properties are important to evaluate the compatibility of tested samples as tissue mimicking materials (TMMs). Common acoustic measurement systems require distilled water as the propagation medium. However, their accuracies are affected by the small change in the medium density and the inaccurate measurement of water temperature. An alternative acoustic pulse echo immersion measurement system for nonporous TMMs is developed in this study. It is developed based on the alternative pulse echo immersion technique (aPEIT) to improve the previous developed system for the noncontact pulse echo immersion technique (PEIT) and specifically designed for the step-shaped nonporous sample. It consists of a pulser/receiver generator, an unfocused transducer, a digital oscilloscope, a temperature controller and a personal computer which are installed with the custom-developed computer program to determine the longitudinal velocity, acoustic impedance, phase velocity and attenuation coefficient of the sample. The precision and accuracy of the developed system are tested for different thickness of sample, temperature of medium, density of medium and center frequency of transducer. The study indicates that developed system for the aPEIT produces the comparable results within 1.16% differences as the previous developed system for the noncontact PEIT, precise results within 6.38% from the average values and accurate results within 0.62% error compared to the reference values. The developed system for the aPEIT offers comparable but more precise results compared to the previous developed system for the noncontact PEIT in measuring the acoustic properties of nonporous TMMs. It can be operated using online and offline analysis modes to measure and differentiate the acoustic properties of specific types of human tissues and TMMs. 2020 Thesis NonPeerReviewed application/pdf en http://eprints.utm.my/id/eprint/101684/1/AnisNazihahMatDaudPFS2020.pdf Mat Daud, Anis Nazihah (2020) Alternative acoustic pulse echo immersion measurement system development for nonporous tissue mimicking materials. PhD thesis, Universiti Teknologi Malaysia. http://dms.library.utm.my:8080/vital/access/manager/Repository/vital:145905
institution Universiti Teknologi Malaysia
building UTM Library
collection Institutional Repository
continent Asia
country Malaysia
content_provider Universiti Teknologi Malaysia
content_source UTM Institutional Repository
url_provider http://eprints.utm.my/
language English
topic QD Chemistry
spellingShingle QD Chemistry
Mat Daud, Anis Nazihah
Alternative acoustic pulse echo immersion measurement system development for nonporous tissue mimicking materials
description Acoustic properties are important to evaluate the compatibility of tested samples as tissue mimicking materials (TMMs). Common acoustic measurement systems require distilled water as the propagation medium. However, their accuracies are affected by the small change in the medium density and the inaccurate measurement of water temperature. An alternative acoustic pulse echo immersion measurement system for nonporous TMMs is developed in this study. It is developed based on the alternative pulse echo immersion technique (aPEIT) to improve the previous developed system for the noncontact pulse echo immersion technique (PEIT) and specifically designed for the step-shaped nonporous sample. It consists of a pulser/receiver generator, an unfocused transducer, a digital oscilloscope, a temperature controller and a personal computer which are installed with the custom-developed computer program to determine the longitudinal velocity, acoustic impedance, phase velocity and attenuation coefficient of the sample. The precision and accuracy of the developed system are tested for different thickness of sample, temperature of medium, density of medium and center frequency of transducer. The study indicates that developed system for the aPEIT produces the comparable results within 1.16% differences as the previous developed system for the noncontact PEIT, precise results within 6.38% from the average values and accurate results within 0.62% error compared to the reference values. The developed system for the aPEIT offers comparable but more precise results compared to the previous developed system for the noncontact PEIT in measuring the acoustic properties of nonporous TMMs. It can be operated using online and offline analysis modes to measure and differentiate the acoustic properties of specific types of human tissues and TMMs.
format Thesis
author Mat Daud, Anis Nazihah
author_facet Mat Daud, Anis Nazihah
author_sort Mat Daud, Anis Nazihah
title Alternative acoustic pulse echo immersion measurement system development for nonporous tissue mimicking materials
title_short Alternative acoustic pulse echo immersion measurement system development for nonporous tissue mimicking materials
title_full Alternative acoustic pulse echo immersion measurement system development for nonporous tissue mimicking materials
title_fullStr Alternative acoustic pulse echo immersion measurement system development for nonporous tissue mimicking materials
title_full_unstemmed Alternative acoustic pulse echo immersion measurement system development for nonporous tissue mimicking materials
title_sort alternative acoustic pulse echo immersion measurement system development for nonporous tissue mimicking materials
publishDate 2020
url http://eprints.utm.my/id/eprint/101684/1/AnisNazihahMatDaudPFS2020.pdf
http://eprints.utm.my/id/eprint/101684/
http://dms.library.utm.my:8080/vital/access/manager/Repository/vital:145905
_version_ 1772811103513870336
score 13.188404