Evaluation of scanner parameters on ebt 2 And ebt 3 film analysis

This study investigates the effects of scanner parameters properties to EBT 2 and EBT 3 films analysis irriadiation with photon and electron beam energy. The scanner parameters evaluated are interpretation (no invert, invert 8-bit scan, invert 16-bit scan), colour to grayscale transformation (Gafc...

Full description

Saved in:
Bibliographic Details
Main Author: Azhan, Siti Nur Amirah
Format: Monograph
Language:English
Published: Universiti Sains Malaysia 2016
Subjects:
Online Access:http://eprints.usm.my/61377/1/SITI%20NUR%20AMIRAH%20BINTI%20AZHAN%20-%20e.pdf
http://eprints.usm.my/61377/
Tags: Add Tag
No Tags, Be the first to tag this record!
id my.usm.eprints.61377
record_format eprints
spelling my.usm.eprints.61377 http://eprints.usm.my/61377/ Evaluation of scanner parameters on ebt 2 And ebt 3 film analysis Azhan, Siti Nur Amirah RA440-440.87 Study and teaching. Research RC71-78.7 Examination. Diagnosis Including raiography This study investigates the effects of scanner parameters properties to EBT 2 and EBT 3 films analysis irriadiation with photon and electron beam energy. The scanner parameters evaluated are interpretation (no invert, invert 8-bit scan, invert 16-bit scan), colour to grayscale transformation (Gafchromic optimised and user-defined transformation), flattening correction (no filter and median filter), film type (positive and colour negative film), image type (bit colour and grayscale), and resolution (72 dpi, 300 dpi, 600 dpi). The EBT 2 and EBT 3 films were initially irradiated with photon and electron beams of different energies (6 MV, 10 MV, 6 MeV and 9 MeV) using Siemens Primus Linear Accelerator. After irriadiation, the films were then scanned using Epson Expression 10000 XL flatbed using different scanning parameters. As a result, the scanner flattening correction did not have any effects the optical density (OD) reading of EBT 2 and EBT 3 films. However, the interpretation section of scanner parameter did changes if invert 16-bit scan was used compared to similar reading obtained if invert 8- bit scan and standard parameter (no invert) were used. The OD of film type is significantly different than the standard parameter and the image type also affect the OD value. The different parameter of resolution flatbed scanner gives almost consistent OD value but changes to the higher resolution will increase the OD value. In conclusion, the scanning parameters have significant effects on the films OD analysis and hence affect the dose measured. Universiti Sains Malaysia 2016 Monograph NonPeerReviewed application/pdf en http://eprints.usm.my/61377/1/SITI%20NUR%20AMIRAH%20BINTI%20AZHAN%20-%20e.pdf Azhan, Siti Nur Amirah (2016) Evaluation of scanner parameters on ebt 2 And ebt 3 film analysis. Project Report. Universiti Sains Malaysia. (Submitted)
institution Universiti Sains Malaysia
building Hamzah Sendut Library
collection Institutional Repository
continent Asia
country Malaysia
content_provider Universiti Sains Malaysia
content_source USM Institutional Repository
url_provider http://eprints.usm.my/
language English
topic RA440-440.87 Study and teaching. Research
RC71-78.7 Examination. Diagnosis Including raiography
spellingShingle RA440-440.87 Study and teaching. Research
RC71-78.7 Examination. Diagnosis Including raiography
Azhan, Siti Nur Amirah
Evaluation of scanner parameters on ebt 2 And ebt 3 film analysis
description This study investigates the effects of scanner parameters properties to EBT 2 and EBT 3 films analysis irriadiation with photon and electron beam energy. The scanner parameters evaluated are interpretation (no invert, invert 8-bit scan, invert 16-bit scan), colour to grayscale transformation (Gafchromic optimised and user-defined transformation), flattening correction (no filter and median filter), film type (positive and colour negative film), image type (bit colour and grayscale), and resolution (72 dpi, 300 dpi, 600 dpi). The EBT 2 and EBT 3 films were initially irradiated with photon and electron beams of different energies (6 MV, 10 MV, 6 MeV and 9 MeV) using Siemens Primus Linear Accelerator. After irriadiation, the films were then scanned using Epson Expression 10000 XL flatbed using different scanning parameters. As a result, the scanner flattening correction did not have any effects the optical density (OD) reading of EBT 2 and EBT 3 films. However, the interpretation section of scanner parameter did changes if invert 16-bit scan was used compared to similar reading obtained if invert 8- bit scan and standard parameter (no invert) were used. The OD of film type is significantly different than the standard parameter and the image type also affect the OD value. The different parameter of resolution flatbed scanner gives almost consistent OD value but changes to the higher resolution will increase the OD value. In conclusion, the scanning parameters have significant effects on the films OD analysis and hence affect the dose measured.
format Monograph
author Azhan, Siti Nur Amirah
author_facet Azhan, Siti Nur Amirah
author_sort Azhan, Siti Nur Amirah
title Evaluation of scanner parameters on ebt 2 And ebt 3 film analysis
title_short Evaluation of scanner parameters on ebt 2 And ebt 3 film analysis
title_full Evaluation of scanner parameters on ebt 2 And ebt 3 film analysis
title_fullStr Evaluation of scanner parameters on ebt 2 And ebt 3 film analysis
title_full_unstemmed Evaluation of scanner parameters on ebt 2 And ebt 3 film analysis
title_sort evaluation of scanner parameters on ebt 2 and ebt 3 film analysis
publisher Universiti Sains Malaysia
publishDate 2016
url http://eprints.usm.my/61377/1/SITI%20NUR%20AMIRAH%20BINTI%20AZHAN%20-%20e.pdf
http://eprints.usm.my/61377/
_version_ 1823273880419041280
score 13.239859