Correlation Between Pull-Out Strength And Insertion Strain Field In Relation To Bone Screw Insertion Angle
In this final year project, an investigation of the correlation between pull-out strength and insertion strain field in relation to bone screw insertion angle is studied. In general, cortical screw is one of the internal fixation devices that surgeons used to perform orthopedic surgery however, stud...
Saved in:
Main Author: | |
---|---|
Format: | Monograph |
Language: | English |
Published: |
Universiti Sains Malaysia
2021
|
Subjects: | |
Online Access: | http://eprints.usm.my/55970/1/Correlation%20Between%20Pull-Out%20Strength%20And%20Insertion%20Strain%20Field%20In%20Relation%20To%20Bone%20Screw%20Insertion%20Angle.pdf http://eprints.usm.my/55970/ |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
id |
my.usm.eprints.55970 |
---|---|
record_format |
eprints |
spelling |
my.usm.eprints.55970 http://eprints.usm.my/55970/ Correlation Between Pull-Out Strength And Insertion Strain Field In Relation To Bone Screw Insertion Angle Zulkifle, Muhammad Syamim T Technology TJ1 Mechanical engineering and machinery In this final year project, an investigation of the correlation between pull-out strength and insertion strain field in relation to bone screw insertion angle is studied. In general, cortical screw is one of the internal fixation devices that surgeons used to perform orthopedic surgery however, studies have found that the use of these devices may cause hardware loosening. Many studies have come out with several solutions which are changing the screw material and alter the screw insertion depth. However, this study will explain on how the screw insertion angle influences the pull-out force of the screw as the strain field developed within the bone, and the affects the pull-out force. A cortical screw and 0.32 g/cm3 rigid PU foam block are the main material for this experiment. There are two experiments conducted which are the Screw Insertion and Screw Pull-out Test that will be done according to the ASTM standards. In screw insertion test, screw insertion strain developed within the PU foam block are analyzed using Digital Image Correlation (DIC) system and Davis 10 software. While in screw pull-out test, the pull-out strength is analyzed and calculated using a specialized software with the UTM Instron 8874. The results show that the maximum recorded screw pull-out force is 607.43 N when the screw is inserted at 30⁰, the highest strain displacement in x and y direction is 0.0115 at 15⁰ screw insertion and 0.025 at 0⁰ screw insertion, respectively. Universiti Sains Malaysia 2021-07-01 Monograph NonPeerReviewed application/pdf en http://eprints.usm.my/55970/1/Correlation%20Between%20Pull-Out%20Strength%20And%20Insertion%20Strain%20Field%20In%20Relation%20To%20Bone%20Screw%20Insertion%20Angle.pdf Zulkifle, Muhammad Syamim (2021) Correlation Between Pull-Out Strength And Insertion Strain Field In Relation To Bone Screw Insertion Angle. Project Report. Universiti Sains Malaysia, Pusat Pengajian Kejuruteraan Mekanik. (Submitted) |
institution |
Universiti Sains Malaysia |
building |
Hamzah Sendut Library |
collection |
Institutional Repository |
continent |
Asia |
country |
Malaysia |
content_provider |
Universiti Sains Malaysia |
content_source |
USM Institutional Repository |
url_provider |
http://eprints.usm.my/ |
language |
English |
topic |
T Technology TJ1 Mechanical engineering and machinery |
spellingShingle |
T Technology TJ1 Mechanical engineering and machinery Zulkifle, Muhammad Syamim Correlation Between Pull-Out Strength And Insertion Strain Field In Relation To Bone Screw Insertion Angle |
description |
In this final year project, an investigation of the correlation between pull-out strength and insertion strain field in relation to bone screw insertion angle is studied. In general, cortical screw is one of the internal fixation devices that surgeons used to perform orthopedic surgery however, studies have found that the use of these devices may cause hardware loosening. Many studies have come out with several solutions which are changing the screw material and alter the screw insertion depth. However, this study will explain on how the screw insertion angle influences the pull-out force of the screw as the strain field developed within the bone, and the affects the pull-out force. A cortical screw and 0.32 g/cm3 rigid PU foam block are the main material for this experiment. There are two experiments conducted which are the Screw Insertion and Screw Pull-out Test that will be done according to the ASTM standards. In screw insertion test, screw insertion strain developed within the PU foam block are analyzed using Digital Image Correlation (DIC) system and Davis 10 software. While in screw pull-out test, the pull-out strength is analyzed and calculated using a specialized software with the UTM Instron 8874. The results show that the maximum recorded screw pull-out force is 607.43 N when the screw is inserted at 30⁰, the highest strain displacement in x and y direction is 0.0115 at 15⁰ screw insertion and 0.025 at 0⁰ screw insertion, respectively. |
format |
Monograph |
author |
Zulkifle, Muhammad Syamim |
author_facet |
Zulkifle, Muhammad Syamim |
author_sort |
Zulkifle, Muhammad Syamim |
title |
Correlation Between Pull-Out Strength And Insertion Strain Field In Relation To Bone Screw Insertion Angle |
title_short |
Correlation Between Pull-Out Strength And Insertion Strain Field In Relation To Bone Screw Insertion Angle |
title_full |
Correlation Between Pull-Out Strength And Insertion Strain Field In Relation To Bone Screw Insertion Angle |
title_fullStr |
Correlation Between Pull-Out Strength And Insertion Strain Field In Relation To Bone Screw Insertion Angle |
title_full_unstemmed |
Correlation Between Pull-Out Strength And Insertion Strain Field In Relation To Bone Screw Insertion Angle |
title_sort |
correlation between pull-out strength and insertion strain field in relation to bone screw insertion angle |
publisher |
Universiti Sains Malaysia |
publishDate |
2021 |
url |
http://eprints.usm.my/55970/1/Correlation%20Between%20Pull-Out%20Strength%20And%20Insertion%20Strain%20Field%20In%20Relation%20To%20Bone%20Screw%20Insertion%20Angle.pdf http://eprints.usm.my/55970/ |
_version_ |
1752147250504531968 |
score |
13.211869 |