A study of continuous monitoring of endotracheal tube cuff pressure using an electronic device in critical care setting

The endotracheal tube (ETT) cuff pressure measurement is not a routine practice in many critical care units. The main purpose of this study is to determine the accuracy or relationship of electronic measurement of ETT cuff pressure by continuous method (using aneroid manometer method as the contr...

Full description

Saved in:
Bibliographic Details
Main Author: Khairuddin, Ismail
Format: Thesis
Language:English
Published: 2009
Subjects:
Online Access:http://eprints.usm.my/53961/1/DR%20KHAIRUDDIN%20BIN%20ISMAIL%20-%2024%20pages.pdf
http://eprints.usm.my/53961/
Tags: Add Tag
No Tags, Be the first to tag this record!
id my.usm.eprints.53961
record_format eprints
spelling my.usm.eprints.53961 http://eprints.usm.my/53961/ A study of continuous monitoring of endotracheal tube cuff pressure using an electronic device in critical care setting Khairuddin, Ismail R Medicine (General) The endotracheal tube (ETT) cuff pressure measurement is not a routine practice in many critical care units. The main purpose of this study is to determine the accuracy or relationship of electronic measurement of ETT cuff pressure by continuous method (using aneroid manometer method as the control) and the usefulness of the continuous method toward reducing endotracheal tube-related complications in intensive care settings. A total of 56 patients were recruited involving ICU patient intubated with ETT with cuff. A preliminary study was also conducted to determine the safety and stability of this electronic device. The precision of reading of ETT cuff pressure were compared with manual aneroid manometer and the variability of cuff pressure changes were also assessed apart from ETT cuff related complications. The complications that we looked for were the prevalence of ventilator-associated pneumonia (V AP), post-extubation stridor, sore-throat and endotrachealtube (ETT) dislodgement. We found that there was a significant correlation between the readings by aneroid manometer with that of electronic measurement device, with p value of< 0.001 ( mean 44.16 vs 44.22 mm Hg ). The prevalence of V AP, sore-throat and endotracheal tube dislodgement were 1.8%, 3.6% and 3.6% respectively. No patient was noticed to have post-extubation stridor. The readings recorded by the electronic device for ETT cuff measurement was comparable with aneroid manometer and its use could potentially reduce ETT cuff-related complications. 2009 Thesis NonPeerReviewed application/pdf en http://eprints.usm.my/53961/1/DR%20KHAIRUDDIN%20BIN%20ISMAIL%20-%2024%20pages.pdf Khairuddin, Ismail (2009) A study of continuous monitoring of endotracheal tube cuff pressure using an electronic device in critical care setting. Masters thesis, Pusat Pengajian Sains Perubatan.
institution Universiti Sains Malaysia
building Hamzah Sendut Library
collection Institutional Repository
continent Asia
country Malaysia
content_provider Universiti Sains Malaysia
content_source USM Institutional Repository
url_provider http://eprints.usm.my/
language English
topic R Medicine (General)
spellingShingle R Medicine (General)
Khairuddin, Ismail
A study of continuous monitoring of endotracheal tube cuff pressure using an electronic device in critical care setting
description The endotracheal tube (ETT) cuff pressure measurement is not a routine practice in many critical care units. The main purpose of this study is to determine the accuracy or relationship of electronic measurement of ETT cuff pressure by continuous method (using aneroid manometer method as the control) and the usefulness of the continuous method toward reducing endotracheal tube-related complications in intensive care settings. A total of 56 patients were recruited involving ICU patient intubated with ETT with cuff. A preliminary study was also conducted to determine the safety and stability of this electronic device. The precision of reading of ETT cuff pressure were compared with manual aneroid manometer and the variability of cuff pressure changes were also assessed apart from ETT cuff related complications. The complications that we looked for were the prevalence of ventilator-associated pneumonia (V AP), post-extubation stridor, sore-throat and endotrachealtube (ETT) dislodgement. We found that there was a significant correlation between the readings by aneroid manometer with that of electronic measurement device, with p value of< 0.001 ( mean 44.16 vs 44.22 mm Hg ). The prevalence of V AP, sore-throat and endotracheal tube dislodgement were 1.8%, 3.6% and 3.6% respectively. No patient was noticed to have post-extubation stridor. The readings recorded by the electronic device for ETT cuff measurement was comparable with aneroid manometer and its use could potentially reduce ETT cuff-related complications.
format Thesis
author Khairuddin, Ismail
author_facet Khairuddin, Ismail
author_sort Khairuddin, Ismail
title A study of continuous monitoring of endotracheal tube cuff pressure using an electronic device in critical care setting
title_short A study of continuous monitoring of endotracheal tube cuff pressure using an electronic device in critical care setting
title_full A study of continuous monitoring of endotracheal tube cuff pressure using an electronic device in critical care setting
title_fullStr A study of continuous monitoring of endotracheal tube cuff pressure using an electronic device in critical care setting
title_full_unstemmed A study of continuous monitoring of endotracheal tube cuff pressure using an electronic device in critical care setting
title_sort study of continuous monitoring of endotracheal tube cuff pressure using an electronic device in critical care setting
publishDate 2009
url http://eprints.usm.my/53961/1/DR%20KHAIRUDDIN%20BIN%20ISMAIL%20-%2024%20pages.pdf
http://eprints.usm.my/53961/
_version_ 1740826600538112000
score 13.214268