Microfluidic paper-based analytical device (μPAD) for rapid detection of cadmium in river water

In recent years, the issue of contamination of water with heavy metal ion is never-ending, thus seeking extensive attention from all over the world. Heavy metal above a threshold limit is toxic to humans as well as the environment. Therefore, a very promising method using μPAD is presented for th...

Full description

Saved in:
Bibliographic Details
Main Author: Aziz, Nor Izati Che Ab
Format: Thesis
Language:English
Published: 2020
Subjects:
Online Access:http://eprints.usm.my/49115/1/TESIS-NOR%20IZATI%20CHE%20AB%20AZIZ%20%28P-SKM003719%29%202-24%20pages.pdf
http://eprints.usm.my/49115/
Tags: Add Tag
No Tags, Be the first to tag this record!
id my.usm.eprints.49115
record_format eprints
spelling my.usm.eprints.49115 http://eprints.usm.my/49115/ Microfluidic paper-based analytical device (μPAD) for rapid detection of cadmium in river water Aziz, Nor Izati Che Ab R Medicine In recent years, the issue of contamination of water with heavy metal ion is never-ending, thus seeking extensive attention from all over the world. Heavy metal above a threshold limit is toxic to humans as well as the environment. Therefore, a very promising method using μPAD is presented for the rapid determination of Cd at an ultra-trace level using 1,5-diphenylthiocarbazone (dithizone) as colorimetric reagent (λ=500 nm) in aqueous solution. The mixture turns from brownish-red to orange colour in the presence of Cd. The use of paraffin wax with the handheld rubber stamp was demonstrated for the detection of Cd using colorimetric detection. The paraffin wax was used onto the filter paper to create a hydrophobic barrier for fluidic channels. The μPAD was fabricated within 10 min and provided high reproducibility and stability. The rubber stamping method provides a simple, rapid and cost-effective in fabrication of uPAD. The calibration curves were constructed for developed method of μPAD. The linear coefficient (R2) was 0.9538 and the detection limits of Cd ion using μPAD was 3.87 ppm. The results demonstrate that the detection limits obtained for both methods were higher than the permissible limit of Cd which is 0.005 ppm. These methods were reported to be less sensitive for ultra-trace metal such as Cd. However, for real sample analysis, the recovery of Cd in the water samples was measured as 82.5% using μPAD. It has good recovery value. To conclude, the fabricated μPAD can be used as screening test which one of the most vital steps in paper-based assays to identify the presence of analyte but confirmation step is required for further analysis using the sophisticated instruments. 2020-09 Thesis NonPeerReviewed application/pdf en http://eprints.usm.my/49115/1/TESIS-NOR%20IZATI%20CHE%20AB%20AZIZ%20%28P-SKM003719%29%202-24%20pages.pdf Aziz, Nor Izati Che Ab (2020) Microfluidic paper-based analytical device (μPAD) for rapid detection of cadmium in river water. Masters thesis, Universiti Sains Malaysia.
institution Universiti Sains Malaysia
building Hamzah Sendut Library
collection Institutional Repository
continent Asia
country Malaysia
content_provider Universiti Sains Malaysia
content_source USM Institutional Repository
url_provider http://eprints.usm.my/
language English
topic R Medicine
spellingShingle R Medicine
Aziz, Nor Izati Che Ab
Microfluidic paper-based analytical device (μPAD) for rapid detection of cadmium in river water
description In recent years, the issue of contamination of water with heavy metal ion is never-ending, thus seeking extensive attention from all over the world. Heavy metal above a threshold limit is toxic to humans as well as the environment. Therefore, a very promising method using μPAD is presented for the rapid determination of Cd at an ultra-trace level using 1,5-diphenylthiocarbazone (dithizone) as colorimetric reagent (λ=500 nm) in aqueous solution. The mixture turns from brownish-red to orange colour in the presence of Cd. The use of paraffin wax with the handheld rubber stamp was demonstrated for the detection of Cd using colorimetric detection. The paraffin wax was used onto the filter paper to create a hydrophobic barrier for fluidic channels. The μPAD was fabricated within 10 min and provided high reproducibility and stability. The rubber stamping method provides a simple, rapid and cost-effective in fabrication of uPAD. The calibration curves were constructed for developed method of μPAD. The linear coefficient (R2) was 0.9538 and the detection limits of Cd ion using μPAD was 3.87 ppm. The results demonstrate that the detection limits obtained for both methods were higher than the permissible limit of Cd which is 0.005 ppm. These methods were reported to be less sensitive for ultra-trace metal such as Cd. However, for real sample analysis, the recovery of Cd in the water samples was measured as 82.5% using μPAD. It has good recovery value. To conclude, the fabricated μPAD can be used as screening test which one of the most vital steps in paper-based assays to identify the presence of analyte but confirmation step is required for further analysis using the sophisticated instruments.
format Thesis
author Aziz, Nor Izati Che Ab
author_facet Aziz, Nor Izati Che Ab
author_sort Aziz, Nor Izati Che Ab
title Microfluidic paper-based analytical device (μPAD) for rapid detection of cadmium in river water
title_short Microfluidic paper-based analytical device (μPAD) for rapid detection of cadmium in river water
title_full Microfluidic paper-based analytical device (μPAD) for rapid detection of cadmium in river water
title_fullStr Microfluidic paper-based analytical device (μPAD) for rapid detection of cadmium in river water
title_full_unstemmed Microfluidic paper-based analytical device (μPAD) for rapid detection of cadmium in river water
title_sort microfluidic paper-based analytical device (μpad) for rapid detection of cadmium in river water
publishDate 2020
url http://eprints.usm.my/49115/1/TESIS-NOR%20IZATI%20CHE%20AB%20AZIZ%20%28P-SKM003719%29%202-24%20pages.pdf
http://eprints.usm.my/49115/
_version_ 1699238021141889024
score 13.188404