Effect of pysical filters (aluminum 0.2mm and o.jmm thick) on sensitivity, uniformity, and spatial resolution of Toshiba GCA- 901A/HG gamma camera with tc-99m in planar imaging

The performance of a scintillation-camera system is assessed to assure the acquisition of diagnostically reliable images. Decisions-making data to determine acceptability of camera performance are acquired when some parameters are tested. So, in this study the sensitivity, uniformity and system s...

Full description

Saved in:
Bibliographic Details
Main Author: Nawi, Siti Nazipah Mohamad
Format: Article
Language:English
Published: Pusat Pengajian Sains Kesihatan, Universiti Sains Malaysia 2004
Subjects:
Online Access:http://eprints.usm.my/46757/1/PTA...Siti%20Nazipah%20Binti%20Mohamad%20Nawi...2004...-24%20pages.pdf
http://eprints.usm.my/46757/
Tags: Add Tag
No Tags, Be the first to tag this record!
id my.usm.eprints.46757
record_format eprints
spelling my.usm.eprints.46757 http://eprints.usm.my/46757/ Effect of pysical filters (aluminum 0.2mm and o.jmm thick) on sensitivity, uniformity, and spatial resolution of Toshiba GCA- 901A/HG gamma camera with tc-99m in planar imaging Nawi, Siti Nazipah Mohamad R Medicine The performance of a scintillation-camera system is assessed to assure the acquisition of diagnostically reliable images. Decisions-making data to determine acceptability of camera performance are acquired when some parameters are tested. So, in this study the sensitivity, uniformity and system spatial resolution parameters are tested by using physical filters, aluminum 0.2 mm and aluminum 0.3 mm. The Tc-99m is used as a source. Two collimator are used which are low energy parallel-hole collimators, i.e. low energy high resolution (L.E.H.R) and low energy general purpose (L.E.G.P.). The reason for this choice was that these collimators are the most commonly used for clinical studies in nuclear medicine departments. For data acquisition, the method utilizes a photopeak energy window ranging 126 keV-154 keV ofTc-99m spectrum with 128 x 128 matrix size were used for all tests. From sensitivity, uniformity, and spatial resolution results material filtered data shows an improvement. Pusat Pengajian Sains Kesihatan, Universiti Sains Malaysia 2004 Article NonPeerReviewed application/pdf en http://eprints.usm.my/46757/1/PTA...Siti%20Nazipah%20Binti%20Mohamad%20Nawi...2004...-24%20pages.pdf Nawi, Siti Nazipah Mohamad (2004) Effect of pysical filters (aluminum 0.2mm and o.jmm thick) on sensitivity, uniformity, and spatial resolution of Toshiba GCA- 901A/HG gamma camera with tc-99m in planar imaging. Effect of pysical filters (aluminum 0.2mm and o.jmm thick) on sensitivity, uniformity, and spatial resolution of Toshiba GCA- 901A/HG gamma camera with tc-99m in planar imaging. (Submitted)
institution Universiti Sains Malaysia
building Hamzah Sendut Library
collection Institutional Repository
continent Asia
country Malaysia
content_provider Universiti Sains Malaysia
content_source USM Institutional Repository
url_provider http://eprints.usm.my/
language English
topic R Medicine
spellingShingle R Medicine
Nawi, Siti Nazipah Mohamad
Effect of pysical filters (aluminum 0.2mm and o.jmm thick) on sensitivity, uniformity, and spatial resolution of Toshiba GCA- 901A/HG gamma camera with tc-99m in planar imaging
description The performance of a scintillation-camera system is assessed to assure the acquisition of diagnostically reliable images. Decisions-making data to determine acceptability of camera performance are acquired when some parameters are tested. So, in this study the sensitivity, uniformity and system spatial resolution parameters are tested by using physical filters, aluminum 0.2 mm and aluminum 0.3 mm. The Tc-99m is used as a source. Two collimator are used which are low energy parallel-hole collimators, i.e. low energy high resolution (L.E.H.R) and low energy general purpose (L.E.G.P.). The reason for this choice was that these collimators are the most commonly used for clinical studies in nuclear medicine departments. For data acquisition, the method utilizes a photopeak energy window ranging 126 keV-154 keV ofTc-99m spectrum with 128 x 128 matrix size were used for all tests. From sensitivity, uniformity, and spatial resolution results material filtered data shows an improvement.
format Article
author Nawi, Siti Nazipah Mohamad
author_facet Nawi, Siti Nazipah Mohamad
author_sort Nawi, Siti Nazipah Mohamad
title Effect of pysical filters (aluminum 0.2mm and o.jmm thick) on sensitivity, uniformity, and spatial resolution of Toshiba GCA- 901A/HG gamma camera with tc-99m in planar imaging
title_short Effect of pysical filters (aluminum 0.2mm and o.jmm thick) on sensitivity, uniformity, and spatial resolution of Toshiba GCA- 901A/HG gamma camera with tc-99m in planar imaging
title_full Effect of pysical filters (aluminum 0.2mm and o.jmm thick) on sensitivity, uniformity, and spatial resolution of Toshiba GCA- 901A/HG gamma camera with tc-99m in planar imaging
title_fullStr Effect of pysical filters (aluminum 0.2mm and o.jmm thick) on sensitivity, uniformity, and spatial resolution of Toshiba GCA- 901A/HG gamma camera with tc-99m in planar imaging
title_full_unstemmed Effect of pysical filters (aluminum 0.2mm and o.jmm thick) on sensitivity, uniformity, and spatial resolution of Toshiba GCA- 901A/HG gamma camera with tc-99m in planar imaging
title_sort effect of pysical filters (aluminum 0.2mm and o.jmm thick) on sensitivity, uniformity, and spatial resolution of toshiba gca- 901a/hg gamma camera with tc-99m in planar imaging
publisher Pusat Pengajian Sains Kesihatan, Universiti Sains Malaysia
publishDate 2004
url http://eprints.usm.my/46757/1/PTA...Siti%20Nazipah%20Binti%20Mohamad%20Nawi...2004...-24%20pages.pdf
http://eprints.usm.my/46757/
_version_ 1674067173166809088