Effect of pysical filters (aluminum 0.2mm and o.jmm thick) on sensitivity, uniformity, and spatial resolution of Toshiba GCA- 901A/HG gamma camera with tc-99m in planar imaging
The performance of a scintillation-camera system is assessed to assure the acquisition of diagnostically reliable images. Decisions-making data to determine acceptability of camera performance are acquired when some parameters are tested. So, in this study the sensitivity, uniformity and system s...
Saved in:
Main Author: | |
---|---|
Format: | Article |
Language: | English |
Published: |
Pusat Pengajian Sains Kesihatan, Universiti Sains Malaysia
2004
|
Subjects: | |
Online Access: | http://eprints.usm.my/46757/1/PTA...Siti%20Nazipah%20Binti%20Mohamad%20Nawi...2004...-24%20pages.pdf http://eprints.usm.my/46757/ |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
id |
my.usm.eprints.46757 |
---|---|
record_format |
eprints |
spelling |
my.usm.eprints.46757 http://eprints.usm.my/46757/ Effect of pysical filters (aluminum 0.2mm and o.jmm thick) on sensitivity, uniformity, and spatial resolution of Toshiba GCA- 901A/HG gamma camera with tc-99m in planar imaging Nawi, Siti Nazipah Mohamad R Medicine The performance of a scintillation-camera system is assessed to assure the acquisition of diagnostically reliable images. Decisions-making data to determine acceptability of camera performance are acquired when some parameters are tested. So, in this study the sensitivity, uniformity and system spatial resolution parameters are tested by using physical filters, aluminum 0.2 mm and aluminum 0.3 mm. The Tc-99m is used as a source. Two collimator are used which are low energy parallel-hole collimators, i.e. low energy high resolution (L.E.H.R) and low energy general purpose (L.E.G.P.). The reason for this choice was that these collimators are the most commonly used for clinical studies in nuclear medicine departments. For data acquisition, the method utilizes a photopeak energy window ranging 126 keV-154 keV ofTc-99m spectrum with 128 x 128 matrix size were used for all tests. From sensitivity, uniformity, and spatial resolution results material filtered data shows an improvement. Pusat Pengajian Sains Kesihatan, Universiti Sains Malaysia 2004 Article NonPeerReviewed application/pdf en http://eprints.usm.my/46757/1/PTA...Siti%20Nazipah%20Binti%20Mohamad%20Nawi...2004...-24%20pages.pdf Nawi, Siti Nazipah Mohamad (2004) Effect of pysical filters (aluminum 0.2mm and o.jmm thick) on sensitivity, uniformity, and spatial resolution of Toshiba GCA- 901A/HG gamma camera with tc-99m in planar imaging. Effect of pysical filters (aluminum 0.2mm and o.jmm thick) on sensitivity, uniformity, and spatial resolution of Toshiba GCA- 901A/HG gamma camera with tc-99m in planar imaging. (Submitted) |
institution |
Universiti Sains Malaysia |
building |
Hamzah Sendut Library |
collection |
Institutional Repository |
continent |
Asia |
country |
Malaysia |
content_provider |
Universiti Sains Malaysia |
content_source |
USM Institutional Repository |
url_provider |
http://eprints.usm.my/ |
language |
English |
topic |
R Medicine |
spellingShingle |
R Medicine Nawi, Siti Nazipah Mohamad Effect of pysical filters (aluminum 0.2mm and o.jmm thick) on sensitivity, uniformity, and spatial resolution of Toshiba GCA- 901A/HG gamma camera with tc-99m in planar imaging |
description |
The performance of a scintillation-camera system is assessed to assure the acquisition
of diagnostically reliable images. Decisions-making data to determine acceptability of
camera performance are acquired when some parameters are tested. So, in this study the
sensitivity, uniformity and system spatial resolution parameters are tested by using physical
filters, aluminum 0.2 mm and aluminum 0.3 mm. The Tc-99m is used as a source. Two
collimator are used which are low energy parallel-hole collimators, i.e. low energy high
resolution (L.E.H.R) and low energy general purpose (L.E.G.P.). The reason for this choice
was that these collimators are the most commonly used for clinical studies in nuclear
medicine departments. For data acquisition, the method utilizes a photopeak energy window
ranging 126 keV-154 keV ofTc-99m spectrum with 128 x 128 matrix size were used for all
tests. From sensitivity, uniformity, and spatial resolution results material filtered data shows
an improvement. |
format |
Article |
author |
Nawi, Siti Nazipah Mohamad |
author_facet |
Nawi, Siti Nazipah Mohamad |
author_sort |
Nawi, Siti Nazipah Mohamad |
title |
Effect of pysical filters (aluminum 0.2mm
and o.jmm thick) on sensitivity, uniformity,
and spatial resolution of Toshiba GCA-
901A/HG gamma camera with tc-99m in
planar imaging |
title_short |
Effect of pysical filters (aluminum 0.2mm
and o.jmm thick) on sensitivity, uniformity,
and spatial resolution of Toshiba GCA-
901A/HG gamma camera with tc-99m in
planar imaging |
title_full |
Effect of pysical filters (aluminum 0.2mm
and o.jmm thick) on sensitivity, uniformity,
and spatial resolution of Toshiba GCA-
901A/HG gamma camera with tc-99m in
planar imaging |
title_fullStr |
Effect of pysical filters (aluminum 0.2mm
and o.jmm thick) on sensitivity, uniformity,
and spatial resolution of Toshiba GCA-
901A/HG gamma camera with tc-99m in
planar imaging |
title_full_unstemmed |
Effect of pysical filters (aluminum 0.2mm
and o.jmm thick) on sensitivity, uniformity,
and spatial resolution of Toshiba GCA-
901A/HG gamma camera with tc-99m in
planar imaging |
title_sort |
effect of pysical filters (aluminum 0.2mm
and o.jmm thick) on sensitivity, uniformity,
and spatial resolution of toshiba gca-
901a/hg gamma camera with tc-99m in
planar imaging |
publisher |
Pusat Pengajian Sains Kesihatan, Universiti Sains Malaysia |
publishDate |
2004 |
url |
http://eprints.usm.my/46757/1/PTA...Siti%20Nazipah%20Binti%20Mohamad%20Nawi...2004...-24%20pages.pdf http://eprints.usm.my/46757/ |
_version_ |
1674067173166809088 |
score |
13.160551 |