The Impact Of Reconstruction Filters On 99mtc Myocardial Spect Imaging

This prospective and retrospective study was undertaken to investigate the effect of filtering on image quality and functional parameters in myocardial SPECT using various reconstruction filters and their associated parameters. In the image quality study (a prospective study), a cardiac phantom syst...

Full description

Saved in:
Bibliographic Details
Main Author: Yusoff, Muhammad Nur Salihin
Format: Thesis
Language:English
Published: 2010
Subjects:
Online Access:http://eprints.usm.my/43032/1/Muhammad_Nur_Salihin_Bin_Yusoff24.pdf
http://eprints.usm.my/43032/
Tags: Add Tag
No Tags, Be the first to tag this record!
id my.usm.eprints.43032
record_format eprints
spelling my.usm.eprints.43032 http://eprints.usm.my/43032/ The Impact Of Reconstruction Filters On 99mtc Myocardial Spect Imaging Yusoff, Muhammad Nur Salihin R5-130.5 General works This prospective and retrospective study was undertaken to investigate the effect of filtering on image quality and functional parameters in myocardial SPECT using various reconstruction filters and their associated parameters. In the image quality study (a prospective study), a cardiac phantom system (with 1.10 cm cold defect in myocardium) for different 99mTc radioactivity concentration as background was used in the SPECT procedure. The target-background concentration ratio (T/B) of 5.3, 4.0, 3.3, 2.7, 2.0, 1.3, and 1.0 were used to acquire SPECT images. Based on the SPECT studies, the calculated background count to target count ratio (BTR) for each T/B was 0.2, 0.4, 0.6, 0.7, 0.8, and 0.9 respectively (T/B = 1.0 was excluded). The image quality was analyzed based on image contrast, signal-to-noise ratio (SNR), defect size accuracy, and total grade (trade-off between contrast, SNR, and defect size accuracy). The results showed that, different filters produce different values of contrast, SNR, and defect size. The increase in cut off frequency and order of filter resulted in the increase in contrast and SNR, but the accuracy of defect size was reduced. Butterworth filter has the ability to balance between contrast, SNR, and defect size accuracy while Gaussian filter is good in producing image with high contrast and SNR. 2010-06 Thesis NonPeerReviewed application/pdf en http://eprints.usm.my/43032/1/Muhammad_Nur_Salihin_Bin_Yusoff24.pdf Yusoff, Muhammad Nur Salihin (2010) The Impact Of Reconstruction Filters On 99mtc Myocardial Spect Imaging. Masters thesis, Universiti Sains Malaysia.
institution Universiti Sains Malaysia
building Hamzah Sendut Library
collection Institutional Repository
continent Asia
country Malaysia
content_provider Universiti Sains Malaysia
content_source USM Institutional Repository
url_provider http://eprints.usm.my/
language English
topic R5-130.5 General works
spellingShingle R5-130.5 General works
Yusoff, Muhammad Nur Salihin
The Impact Of Reconstruction Filters On 99mtc Myocardial Spect Imaging
description This prospective and retrospective study was undertaken to investigate the effect of filtering on image quality and functional parameters in myocardial SPECT using various reconstruction filters and their associated parameters. In the image quality study (a prospective study), a cardiac phantom system (with 1.10 cm cold defect in myocardium) for different 99mTc radioactivity concentration as background was used in the SPECT procedure. The target-background concentration ratio (T/B) of 5.3, 4.0, 3.3, 2.7, 2.0, 1.3, and 1.0 were used to acquire SPECT images. Based on the SPECT studies, the calculated background count to target count ratio (BTR) for each T/B was 0.2, 0.4, 0.6, 0.7, 0.8, and 0.9 respectively (T/B = 1.0 was excluded). The image quality was analyzed based on image contrast, signal-to-noise ratio (SNR), defect size accuracy, and total grade (trade-off between contrast, SNR, and defect size accuracy). The results showed that, different filters produce different values of contrast, SNR, and defect size. The increase in cut off frequency and order of filter resulted in the increase in contrast and SNR, but the accuracy of defect size was reduced. Butterworth filter has the ability to balance between contrast, SNR, and defect size accuracy while Gaussian filter is good in producing image with high contrast and SNR.
format Thesis
author Yusoff, Muhammad Nur Salihin
author_facet Yusoff, Muhammad Nur Salihin
author_sort Yusoff, Muhammad Nur Salihin
title The Impact Of Reconstruction Filters On 99mtc Myocardial Spect Imaging
title_short The Impact Of Reconstruction Filters On 99mtc Myocardial Spect Imaging
title_full The Impact Of Reconstruction Filters On 99mtc Myocardial Spect Imaging
title_fullStr The Impact Of Reconstruction Filters On 99mtc Myocardial Spect Imaging
title_full_unstemmed The Impact Of Reconstruction Filters On 99mtc Myocardial Spect Imaging
title_sort impact of reconstruction filters on 99mtc myocardial spect imaging
publishDate 2010
url http://eprints.usm.my/43032/1/Muhammad_Nur_Salihin_Bin_Yusoff24.pdf
http://eprints.usm.my/43032/
_version_ 1643710644975828992
score 13.211869