Design and calibration of pinch force measurement using strain gauge for post-stroke patients

Two fingers strength is an indicative measurement of pinch impairment. Conventionally, Fugl Meyer Upper Extremity Assessment (FMA-UE) is the primary standard to measure pinch strength of post-stroke survivors. In literature, the evaluation method performed by the th...

Full description

Saved in:
Bibliographic Details
Main Authors: Alsayed, Abdallah, Raja Ahmad, Raja Mohd Kamil, Ramli, Hafiz Rashidi, As'arry, Azizan
Format: Article
Language:English
Published: Penerbit UTHM 2019
Online Access:http://psasir.upm.edu.my/id/eprint/81098/1/PINCH%20FORCE.pdf
http://psasir.upm.edu.my/id/eprint/81098/
https://publisher.uthm.edu.my/ojs/index.php/ijie
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Two fingers strength is an indicative measurement of pinch impairment. Conventionally, Fugl Meyer Upper Extremity Assessment (FMA-UE) is the primary standard to measure pinch strength of post-stroke survivors. In literature, the evaluation method performed by the therapist is subjective and exposed to inter-rater and intra-rater reliabilities. Recently, force-sensing resistors were implemented to measure two fingers force, but these sensors are subjected to nonlinearity, high hysteresis, and voltage drift. This paper presents a design of pinch force measurement based on the strain gauge. The pinch sensor was calibrated within a range of between 0 N to 50 N over a pinching length of 20 mm with a linearity error of 0.0123% and hysteresis of 0.513% during the loading and unloading process. The voltage drift has an average of 0.24% over 20 minutes. The pinch force measurement system reveals an objective pinch force measurements in evaluating the rehabilitation progress of post-stroke patients.