Dose mapping in Cesium-137 blood irradiator using novel Ge-doped silica optical fibres in comparison with gafchromic EBT-XD film: a preliminary study
The purpose of dose-mapping in blood irradiation is to ensure radiation dose delivered to the blood bags is within the specified limits of minimum 15 Gy and maximum 50 Gy, thus preventing any blood transfusion-related diseases. In the present study, we investigated the feasibility of novel fabricate...
Saved in:
Main Authors: | , , , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Elsevier
2023
|
Online Access: | http://psasir.upm.edu.my/id/eprint/107406/1/107406.pdf http://psasir.upm.edu.my/id/eprint/107406/ https://www.sciencedirect.com/science/article/pii/S0969806X23002554?via%3Dihub |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
id |
my.upm.eprints.107406 |
---|---|
record_format |
eprints |
spelling |
my.upm.eprints.1074062024-11-04T03:08:04Z http://psasir.upm.edu.my/id/eprint/107406/ Dose mapping in Cesium-137 blood irradiator using novel Ge-doped silica optical fibres in comparison with gafchromic EBT-XD film: a preliminary study Bakar, K.S.A.K. Noor, N.M. Hassan, M.F. Idris, F. Bradley, D.A. The purpose of dose-mapping in blood irradiation is to ensure radiation dose delivered to the blood bags is within the specified limits of minimum 15 Gy and maximum 50 Gy, thus preventing any blood transfusion-related diseases. In the present study, we investigated the feasibility of novel fabricated germanium-doped (Ge-doped) silica-based optical fibre as a potential passive dosimeter for dose-mapping of a Cesium-137 (Cs-137) gamma blood irradiator. The calibration curve of optical fibres and Gafchromic EBT-XD film was established by irradiating the samples in the dose range from 5 to 35 Gy at the central of a water-equivalent phantom in a dedicated blood irradiator machine. For dose mapping, the fabricated optical fibres and EBT-XD film were irradiated for a gamma exposure of 8 min and 56 s to deliver a central dose of 25 Gy. Thermoluminescence measurements were performed using a Harshaw 3500 TLD reader at a maximum acquisition temperature of 400 °C. The results of optical fibres were compared with EBT-XD film. The absorbed dose obtained from optical fibres shows deviation of within 0.1“10.4 compared to that of dose of the film. The fabricated optical fibre studied in this work have a good potential as a new passive dosimeter for characterisation and dose mapping of gamma blood irradiator due to its high sensitivity and reproducible characteristics. Elsevier 2023-08 Article PeerReviewed text en http://psasir.upm.edu.my/id/eprint/107406/1/107406.pdf Bakar, K.S.A.K. and Noor, N.M. and Hassan, M.F. and Idris, F. and Bradley, D.A. (2023) Dose mapping in Cesium-137 blood irradiator using novel Ge-doped silica optical fibres in comparison with gafchromic EBT-XD film: a preliminary study. Radiation Physics and Chemistry, 209. art. no. 111010. pp. 1-7. ISSN 0969-806X; eISSN: 1879-0895 https://www.sciencedirect.com/science/article/pii/S0969806X23002554?via%3Dihub 10.1016/j.radphyschem.2023.111010 |
institution |
Universiti Putra Malaysia |
building |
UPM Library |
collection |
Institutional Repository |
continent |
Asia |
country |
Malaysia |
content_provider |
Universiti Putra Malaysia |
content_source |
UPM Institutional Repository |
url_provider |
http://psasir.upm.edu.my/ |
language |
English |
description |
The purpose of dose-mapping in blood irradiation is to ensure radiation dose delivered to the blood bags is within the specified limits of minimum 15 Gy and maximum 50 Gy, thus preventing any blood transfusion-related diseases. In the present study, we investigated the feasibility of novel fabricated germanium-doped (Ge-doped) silica-based optical fibre as a potential passive dosimeter for dose-mapping of a Cesium-137 (Cs-137) gamma blood irradiator. The calibration curve of optical fibres and Gafchromic EBT-XD film was established by irradiating the samples in the dose range from 5 to 35 Gy at the central of a water-equivalent phantom in a dedicated blood irradiator machine. For dose mapping, the fabricated optical fibres and EBT-XD film were irradiated for a gamma exposure of 8 min and 56 s to deliver a central dose of 25 Gy. Thermoluminescence measurements were performed using a Harshaw 3500 TLD reader at a maximum acquisition temperature of 400 °C. The results of optical fibres were compared with EBT-XD film. The absorbed dose obtained from optical fibres shows deviation of within 0.1“10.4 compared to that of dose of the film. The fabricated optical fibre studied in this work have a good potential as a new passive dosimeter for characterisation and dose mapping of gamma blood irradiator due to its high sensitivity and reproducible characteristics. |
format |
Article |
author |
Bakar, K.S.A.K. Noor, N.M. Hassan, M.F. Idris, F. Bradley, D.A. |
spellingShingle |
Bakar, K.S.A.K. Noor, N.M. Hassan, M.F. Idris, F. Bradley, D.A. Dose mapping in Cesium-137 blood irradiator using novel Ge-doped silica optical fibres in comparison with gafchromic EBT-XD film: a preliminary study |
author_facet |
Bakar, K.S.A.K. Noor, N.M. Hassan, M.F. Idris, F. Bradley, D.A. |
author_sort |
Bakar, K.S.A.K. |
title |
Dose mapping in Cesium-137 blood irradiator using novel Ge-doped silica optical fibres in comparison with gafchromic EBT-XD film: a preliminary study |
title_short |
Dose mapping in Cesium-137 blood irradiator using novel Ge-doped silica optical fibres in comparison with gafchromic EBT-XD film: a preliminary study |
title_full |
Dose mapping in Cesium-137 blood irradiator using novel Ge-doped silica optical fibres in comparison with gafchromic EBT-XD film: a preliminary study |
title_fullStr |
Dose mapping in Cesium-137 blood irradiator using novel Ge-doped silica optical fibres in comparison with gafchromic EBT-XD film: a preliminary study |
title_full_unstemmed |
Dose mapping in Cesium-137 blood irradiator using novel Ge-doped silica optical fibres in comparison with gafchromic EBT-XD film: a preliminary study |
title_sort |
dose mapping in cesium-137 blood irradiator using novel ge-doped silica optical fibres in comparison with gafchromic ebt-xd film: a preliminary study |
publisher |
Elsevier |
publishDate |
2023 |
url |
http://psasir.upm.edu.my/id/eprint/107406/1/107406.pdf http://psasir.upm.edu.my/id/eprint/107406/ https://www.sciencedirect.com/science/article/pii/S0969806X23002554?via%3Dihub |
_version_ |
1814936553262153728 |
score |
13.211869 |