In house electrostatic field meter calibration for improved ESD protection / Husna Abdul Rahim

This research project presented an improved in house calibration procedure for electrostatic field meter. The proposed improved arrangement is by using copper parallel plates with the optimum separation distance of 6.01 mm between plates. The material properties test, repeatability test and reproduc...