The preparation of in vivo lead x-ray fluorescence system using 99mTc

99mTc is widely used in hospital for nuc1ear medical examination such as scintigraphy of brain and thyroid. The gamma energy emitted by the 99mTc (140.50 keY) is used to produce X-ray fluorescence of the lead atoms present in bones. Being an internal source it is expected the fluorescence efficiency...

Full description

Saved in:
Bibliographic Details
Main Authors: Amin, Yusoff Mohd, Mahat, R.H., Loon, W.M.
Format: Conference or Workshop Item
Language:English
Published: 2005
Subjects:
Online Access:http://eprints.um.edu.my/25043/1/ThePreparation%20of%20InVivo%20LeadX-ray%20Fluorescence%20System%20Using%2099m%20Tc.pdf
http://eprints.um.edu.my/25043/
https://www.osaka-u.ac.jp/en/international/action/network/forum
Tags: Add Tag
No Tags, Be the first to tag this record!
id my.um.eprints.25043
record_format eprints
spelling my.um.eprints.250432020-07-13T01:02:08Z http://eprints.um.edu.my/25043/ The preparation of in vivo lead x-ray fluorescence system using 99mTc Amin, Yusoff Mohd Mahat, R.H. Loon, W.M. Q Science (General) QC Physics 99mTc is widely used in hospital for nuc1ear medical examination such as scintigraphy of brain and thyroid. The gamma energy emitted by the 99mTc (140.50 keY) is used to produce X-ray fluorescence of the lead atoms present in bones. Being an internal source it is expected the fluorescence efficiency to increase. Using a surface barrier X-ray detection system, we have found the minimum detectable concentration (MDC), of lead in bone phantom, is 500 ppm using K-edge X-ray. The bone phantom consists of 10 ml polythene test-tube filled with saline solution and sodium (99mTc) pertechnetate of activity ranging from 7.4 III MBq. The counting time varies from 1-2 hour depending on the activity and lead concentration. 2005 Conference or Workshop Item PeerReviewed text en http://eprints.um.edu.my/25043/1/ThePreparation%20of%20InVivo%20LeadX-ray%20Fluorescence%20System%20Using%2099m%20Tc.pdf Amin, Yusoff Mohd and Mahat, R.H. and Loon, W.M. (2005) The preparation of in vivo lead x-ray fluorescence system using 99mTc. In: Osaka University-Asia Pacific Vietnam National Forum 2005 on Frontiers of Basic Science: Towards New Physics-Earth and Space Science Mathematics, 27-29 September 2005, Hanoi, Vietnam. https://www.osaka-u.ac.jp/en/international/action/network/forum
institution Universiti Malaya
building UM Library
collection Institutional Repository
continent Asia
country Malaysia
content_provider Universiti Malaya
content_source UM Research Repository
url_provider http://eprints.um.edu.my/
language English
topic Q Science (General)
QC Physics
spellingShingle Q Science (General)
QC Physics
Amin, Yusoff Mohd
Mahat, R.H.
Loon, W.M.
The preparation of in vivo lead x-ray fluorescence system using 99mTc
description 99mTc is widely used in hospital for nuc1ear medical examination such as scintigraphy of brain and thyroid. The gamma energy emitted by the 99mTc (140.50 keY) is used to produce X-ray fluorescence of the lead atoms present in bones. Being an internal source it is expected the fluorescence efficiency to increase. Using a surface barrier X-ray detection system, we have found the minimum detectable concentration (MDC), of lead in bone phantom, is 500 ppm using K-edge X-ray. The bone phantom consists of 10 ml polythene test-tube filled with saline solution and sodium (99mTc) pertechnetate of activity ranging from 7.4 III MBq. The counting time varies from 1-2 hour depending on the activity and lead concentration.
format Conference or Workshop Item
author Amin, Yusoff Mohd
Mahat, R.H.
Loon, W.M.
author_facet Amin, Yusoff Mohd
Mahat, R.H.
Loon, W.M.
author_sort Amin, Yusoff Mohd
title The preparation of in vivo lead x-ray fluorescence system using 99mTc
title_short The preparation of in vivo lead x-ray fluorescence system using 99mTc
title_full The preparation of in vivo lead x-ray fluorescence system using 99mTc
title_fullStr The preparation of in vivo lead x-ray fluorescence system using 99mTc
title_full_unstemmed The preparation of in vivo lead x-ray fluorescence system using 99mTc
title_sort preparation of in vivo lead x-ray fluorescence system using 99mtc
publishDate 2005
url http://eprints.um.edu.my/25043/1/ThePreparation%20of%20InVivo%20LeadX-ray%20Fluorescence%20System%20Using%2099m%20Tc.pdf
http://eprints.um.edu.my/25043/
https://www.osaka-u.ac.jp/en/international/action/network/forum
_version_ 1680856995813392384
score 13.18916