Characterization of MOSkin detector for in vivo dose verification during Cobalt-60 high dose-rate intracavitary brachytherapy

In vivo dosimetry in high dose-rate (HDR) intracavitary brachytherapy (ICBT) is important for assessing the true dose received by surrounding organs at risk during treatment. It also serves as part of the treatment delivery quality assurance and verification program with the use of a suitable dosime...

Full description

Saved in:
Bibliographic Details
Main Authors: Jamalludin, Zulaikha, Jong, Wei Loong, Malik, Rozita Abdul, Rosenfeld, Anatoly B., Ung, Ngie Min
Format: Article
Published: Elsevier 2019
Subjects:
Online Access:http://eprints.um.edu.my/24124/
https://doi.org/10.1016/j.ejmp.2019.01.010
Tags: Add Tag
No Tags, Be the first to tag this record!
id my.um.eprints.24124
record_format eprints
spelling my.um.eprints.241242020-03-31T16:16:09Z http://eprints.um.edu.my/24124/ Characterization of MOSkin detector for in vivo dose verification during Cobalt-60 high dose-rate intracavitary brachytherapy Jamalludin, Zulaikha Jong, Wei Loong Malik, Rozita Abdul Rosenfeld, Anatoly B. Ung, Ngie Min R Medicine In vivo dosimetry in high dose-rate (HDR) intracavitary brachytherapy (ICBT) is important for assessing the true dose received by surrounding organs at risk during treatment. It also serves as part of the treatment delivery quality assurance and verification program with the use of a suitable dosimeter. Such a dosimeter should be characterized under brachytherapy conditions before clinical application to ensure the accuracy of in vivo measurement. In this study, a MOSFET-based detector, MOSkin, was calibrated and characterized under HDR Cobalt-60 (Co-60) brachytherapy source. MOSkin possessed the major advantages of having small physical and dosimetric sizes of 4.8 × 10 −6 mm 3 with the ability to provide real-time measurements. Using solid water and polymethyl methacrylate (PMMA) phantom, the detectors’ reproducibility, linearity, angular and distance dependency was tested for its suitability as an in vivo detector. Correction factors to account for differences in depth measurements were determined. The MOSkin detector showed a reliable response when tested under Co-60 brachytherapy range of doses with an excellent linearity of R 2 = 0.9997 and acceptable reproducibility. A phantom verification study was also conducted to verify the differences between MOSkin responses and treatment planning (TPS) calculated doses. By taking into account several correction factors, deviations ranging between 0.01 and 0.4 Gy were found between MOSkin measured and TPS doses at measurement distance of 20–55 mm. The use of MOSkin as the dosimeter of choice for in vivo dosimetry under Co-60 brachytherapy condition is feasible. © 2019 Associazione Italiana di Fisica Medica Elsevier 2019 Article PeerReviewed Jamalludin, Zulaikha and Jong, Wei Loong and Malik, Rozita Abdul and Rosenfeld, Anatoly B. and Ung, Ngie Min (2019) Characterization of MOSkin detector for in vivo dose verification during Cobalt-60 high dose-rate intracavitary brachytherapy. Physica Medica, 58. pp. 1-7. ISSN 1120-1797 https://doi.org/10.1016/j.ejmp.2019.01.010 doi:10.1016/j.ejmp.2019.01.010
institution Universiti Malaya
building UM Library
collection Institutional Repository
continent Asia
country Malaysia
content_provider Universiti Malaya
content_source UM Research Repository
url_provider http://eprints.um.edu.my/
topic R Medicine
spellingShingle R Medicine
Jamalludin, Zulaikha
Jong, Wei Loong
Malik, Rozita Abdul
Rosenfeld, Anatoly B.
Ung, Ngie Min
Characterization of MOSkin detector for in vivo dose verification during Cobalt-60 high dose-rate intracavitary brachytherapy
description In vivo dosimetry in high dose-rate (HDR) intracavitary brachytherapy (ICBT) is important for assessing the true dose received by surrounding organs at risk during treatment. It also serves as part of the treatment delivery quality assurance and verification program with the use of a suitable dosimeter. Such a dosimeter should be characterized under brachytherapy conditions before clinical application to ensure the accuracy of in vivo measurement. In this study, a MOSFET-based detector, MOSkin, was calibrated and characterized under HDR Cobalt-60 (Co-60) brachytherapy source. MOSkin possessed the major advantages of having small physical and dosimetric sizes of 4.8 × 10 −6 mm 3 with the ability to provide real-time measurements. Using solid water and polymethyl methacrylate (PMMA) phantom, the detectors’ reproducibility, linearity, angular and distance dependency was tested for its suitability as an in vivo detector. Correction factors to account for differences in depth measurements were determined. The MOSkin detector showed a reliable response when tested under Co-60 brachytherapy range of doses with an excellent linearity of R 2 = 0.9997 and acceptable reproducibility. A phantom verification study was also conducted to verify the differences between MOSkin responses and treatment planning (TPS) calculated doses. By taking into account several correction factors, deviations ranging between 0.01 and 0.4 Gy were found between MOSkin measured and TPS doses at measurement distance of 20–55 mm. The use of MOSkin as the dosimeter of choice for in vivo dosimetry under Co-60 brachytherapy condition is feasible. © 2019 Associazione Italiana di Fisica Medica
format Article
author Jamalludin, Zulaikha
Jong, Wei Loong
Malik, Rozita Abdul
Rosenfeld, Anatoly B.
Ung, Ngie Min
author_facet Jamalludin, Zulaikha
Jong, Wei Loong
Malik, Rozita Abdul
Rosenfeld, Anatoly B.
Ung, Ngie Min
author_sort Jamalludin, Zulaikha
title Characterization of MOSkin detector for in vivo dose verification during Cobalt-60 high dose-rate intracavitary brachytherapy
title_short Characterization of MOSkin detector for in vivo dose verification during Cobalt-60 high dose-rate intracavitary brachytherapy
title_full Characterization of MOSkin detector for in vivo dose verification during Cobalt-60 high dose-rate intracavitary brachytherapy
title_fullStr Characterization of MOSkin detector for in vivo dose verification during Cobalt-60 high dose-rate intracavitary brachytherapy
title_full_unstemmed Characterization of MOSkin detector for in vivo dose verification during Cobalt-60 high dose-rate intracavitary brachytherapy
title_sort characterization of moskin detector for in vivo dose verification during cobalt-60 high dose-rate intracavitary brachytherapy
publisher Elsevier
publishDate 2019
url http://eprints.um.edu.my/24124/
https://doi.org/10.1016/j.ejmp.2019.01.010
_version_ 1662755226310934528
score 13.188404