A review of surgical and audiological outcomes of bonebridge at tertiary centres in Malaysia
Objectives: To investigate the surgical and audiological outcome of Bonebridge (BB) at tertiary centres in Malaysia. Study Design: Prospective, intra-subject repeated measurements of which each subject is his/her own control, from year 2012 to 2016 at two tertiary referral centres. Methods: Twenty p...
Saved in:
Main Authors: | , , |
---|---|
Format: | Article |
Published: |
Malaysian Medical Association
2018
|
Subjects: | |
Online Access: | http://eprints.um.edu.my/20513/ http://www.e-mjm.org/2018/v73n5/bonebridge.pdf |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
id |
my.um.eprints.20513 |
---|---|
record_format |
eprints |
spelling |
my.um.eprints.205132019-02-27T01:38:53Z http://eprints.um.edu.my/20513/ A review of surgical and audiological outcomes of bonebridge at tertiary centres in Malaysia Tang, Ing Ping Ngui, Ling Xiu Narayanan, Prepageran R Medicine Objectives: To investigate the surgical and audiological outcome of Bonebridge (BB) at tertiary centres in Malaysia. Study Design: Prospective, intra-subject repeated measurements of which each subject is his/her own control, from year 2012 to 2016 at two tertiary referral centres. Methods: Twenty patients with hearing loss who fulfilled criteria for BB and showed good response to bone conduction hearing aid trial were included. Implantations of BB were carried out under general anaesthesia with preoperative computed tomography (CT) planning. Complications were monitored up to six months postoperatively. Subjects’ audiometric thresholds for air conduction, bone conduction and sound field at frequencies of 250Hz to 8kHz were assessed preoperatively and at six months postoperatively. Subjects’ satisfaction was evaluated at 6 months post operatively with Hearing Device Satisfaction Scale (HDSS) questionnaire. Results: There was no major complication reported. Mean aided sound field thresholds showed significant improvement for more than 30dB from 500 to 4000kHz (p<0.05). There was no significant change in mean unaided air conduction and bone conduction thresholds pre and post operatively from 500 to 4000kHz, with a difference of less than 5dB (p>0.05). All the patients were very satisfied (>80%) with the implant, attributing to the promising functional outcome and acceptable cosmetic appearance. Conclusions: BB implantation surgery is safe and is effective in restoring hearing deficits among patients aged five and above with conductive or mixed hearing loss and single-sided hearing loss. Malaysian Medical Association 2018-10 Article PeerReviewed Tang, Ing Ping and Ngui, Ling Xiu and Narayanan, Prepageran (2018) A review of surgical and audiological outcomes of bonebridge at tertiary centres in Malaysia. Medical Journal of Malaysia, 73 (5). pp. 276-280. ISSN 0300-5283 http://www.e-mjm.org/2018/v73n5/bonebridge.pdf |
institution |
Universiti Malaya |
building |
UM Library |
collection |
Institutional Repository |
continent |
Asia |
country |
Malaysia |
content_provider |
Universiti Malaya |
content_source |
UM Research Repository |
url_provider |
http://eprints.um.edu.my/ |
topic |
R Medicine |
spellingShingle |
R Medicine Tang, Ing Ping Ngui, Ling Xiu Narayanan, Prepageran A review of surgical and audiological outcomes of bonebridge at tertiary centres in Malaysia |
description |
Objectives: To investigate the surgical and audiological outcome of Bonebridge (BB) at tertiary centres in Malaysia. Study Design: Prospective, intra-subject repeated measurements of which each subject is his/her own control, from year 2012 to 2016 at two tertiary referral centres. Methods: Twenty patients with hearing loss who fulfilled criteria for BB and showed good response to bone conduction hearing aid trial were included. Implantations of BB were carried out under general anaesthesia with preoperative computed tomography (CT) planning. Complications were monitored up to six months postoperatively. Subjects’ audiometric thresholds for air conduction, bone conduction and sound field at frequencies of 250Hz to 8kHz were assessed preoperatively and at six months postoperatively. Subjects’ satisfaction was evaluated at 6 months post operatively with Hearing Device Satisfaction Scale (HDSS) questionnaire. Results: There was no major complication reported. Mean aided sound field thresholds showed significant improvement for more than 30dB from 500 to 4000kHz (p<0.05). There was no significant change in mean unaided air conduction and bone conduction thresholds pre and post operatively from 500 to 4000kHz, with a difference of less than 5dB (p>0.05). All the patients were very satisfied (>80%) with the implant, attributing to the promising functional outcome and acceptable cosmetic appearance. Conclusions: BB implantation surgery is safe and is effective in restoring hearing deficits among patients aged five and above with conductive or mixed hearing loss and single-sided hearing loss. |
format |
Article |
author |
Tang, Ing Ping Ngui, Ling Xiu Narayanan, Prepageran |
author_facet |
Tang, Ing Ping Ngui, Ling Xiu Narayanan, Prepageran |
author_sort |
Tang, Ing Ping |
title |
A review of surgical and audiological outcomes of bonebridge at tertiary centres in Malaysia |
title_short |
A review of surgical and audiological outcomes of bonebridge at tertiary centres in Malaysia |
title_full |
A review of surgical and audiological outcomes of bonebridge at tertiary centres in Malaysia |
title_fullStr |
A review of surgical and audiological outcomes of bonebridge at tertiary centres in Malaysia |
title_full_unstemmed |
A review of surgical and audiological outcomes of bonebridge at tertiary centres in Malaysia |
title_sort |
review of surgical and audiological outcomes of bonebridge at tertiary centres in malaysia |
publisher |
Malaysian Medical Association |
publishDate |
2018 |
url |
http://eprints.um.edu.my/20513/ http://www.e-mjm.org/2018/v73n5/bonebridge.pdf |
_version_ |
1643691301237948416 |
score |
13.211869 |