Determination of heavy metals in air particulate matter by ion chromatography / Mohd Zahari Abdullah @ Rafie and Rahmalan Ahmad

In recent years, ion chromatography has been increasingly used for the separation and determination of heavy metals. In this study, the separation of Cu2+ Ni2+, Zn2+ and Pb2+ in airborne particuIate sampIes was achieved by the use a complexing eluent containing 20 mM oxalic acid and 20 mM citric aci...

Full description

Saved in:
Bibliographic Details
Main Authors: Abdullah @ Rafie, Mohd Zahari, Ahmad, Rahmalan
Format: Article
Language:English
Published: Universiti Teknologi MARA Cawangan Pahang 2000
Subjects:
Online Access:https://ir.uitm.edu.my/id/eprint/62086/1/62086.PDF
https://ir.uitm.edu.my/id/eprint/62086/
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:In recent years, ion chromatography has been increasingly used for the separation and determination of heavy metals. In this study, the separation of Cu2+ Ni2+, Zn2+ and Pb2+ in airborne particuIate sampIes was achieved by the use a complexing eluent containing 20 mM oxalic acid and 20 mM citric acid adjusted to pH 3.6 with lithium hydroxide. Upon elution, the separated metal ions were reacted with PAR (4-(2-pyridylazo) resorcinol to form coloured complexes in a postcolumn reactor and detected by using a UV-Vis detector at 520 nm. Detection limits for a 25-ul sample solution were 2.2, 0.9, 6.6 and 1.8 ug/l for respectively metals. The technique was optimized and validated by analysing the NIST standard reference materials SRM 1648 Urban Particulate Matter and SRM 1633b Coal Fly Ash.