Design for additive manufacturing and finite element analysis for high flexion total knee replacement (TKR) / Solehuddin Shuib …[et al.]

The patient from the Asian region always demanded a fully functional knee implant, which implied a high-flexion range of motion. Most of their daily life activities utilized the deep knee flexion which flexed until 165° such as the Japanese proper sitting style and Muslim prayer position. The proble...

Full description

Saved in:
Bibliographic Details
Main Authors: Shuib, Solehuddin, Azemi, Mohammad Arsyad, Mohd Arrif, Iffa, Hamizan, Najwa Syakirah
Format: Article
Language:English
Published: Universiti Teknologi MARA 2021
Subjects:
Online Access:http://ir.uitm.edu.my/id/eprint/47679/1/47679.pdf
http://ir.uitm.edu.my/id/eprint/47679/
https://jmeche.uitm.edu.my/
Tags: Add Tag
No Tags, Be the first to tag this record!
id my.uitm.ir.47679
record_format eprints
spelling my.uitm.ir.476792021-06-16T08:49:15Z http://ir.uitm.edu.my/id/eprint/47679/ Design for additive manufacturing and finite element analysis for high flexion total knee replacement (TKR) / Solehuddin Shuib …[et al.] Shuib, Solehuddin Azemi, Mohammad Arsyad Mohd Arrif, Iffa Hamizan, Najwa Syakirah Finite element method Manufacturing processes. Lean manufacturing The patient from the Asian region always demanded a fully functional knee implant, which implied a high-flexion range of motion. Most of their daily life activities utilized the deep knee flexion which flexed until 165° such as the Japanese proper sitting style and Muslim prayer position. The problem of the study is extending the range of motion or achieving the high flexion of total knee replacement as the traditional total knee replacement was incapable to achieve more than 115°. Hence, the purpose of this study is to achieve a modified design of a knee implant that can flex up to 165° by carried out a static structural analysis in the ANSYS R16. There are 0°, 90°, 135°, and 165° angles of flexion with a different net force based on the percentage of body weight implemented on the knee implant. The analysis includes total deformation, Von Mises stress, shear stress, and contact pressure on knee implant were observed and compared to find better modification design. The total deformation had been decreased by about 69% at 0° flexion, 58.5% at 90° flexion, 90.93% at 135° flexion. The contact pressure also had been decreased by about 99.2%, 22.2%, 99.98% at angle flexion of 0, 90, and 135, respectively. The same declination happened to von Mises stress at about 85.05%, 9.52%, and 88.04% at the same angle of 0, 90, and 135, respectively. Universiti Teknologi MARA 2021-04 Article PeerReviewed text en http://ir.uitm.edu.my/id/eprint/47679/1/47679.pdf ID47679 Shuib, Solehuddin and Azemi, Mohammad Arsyad and Mohd Arrif, Iffa and Hamizan, Najwa Syakirah (2021) Design for additive manufacturing and finite element analysis for high flexion total knee replacement (TKR) / Solehuddin Shuib …[et al.]. Journal of Mechanical Engineering (JMechE), 8 (2). pp. 97-110. ISSN (eISSN):2550-164X https://jmeche.uitm.edu.my/
institution Universiti Teknologi Mara
building Tun Abdul Razak Library
collection Institutional Repository
continent Asia
country Malaysia
content_provider Universiti Teknologi Mara
content_source UiTM Institutional Repository
url_provider http://ir.uitm.edu.my/
language English
topic Finite element method
Manufacturing processes. Lean manufacturing
spellingShingle Finite element method
Manufacturing processes. Lean manufacturing
Shuib, Solehuddin
Azemi, Mohammad Arsyad
Mohd Arrif, Iffa
Hamizan, Najwa Syakirah
Design for additive manufacturing and finite element analysis for high flexion total knee replacement (TKR) / Solehuddin Shuib …[et al.]
description The patient from the Asian region always demanded a fully functional knee implant, which implied a high-flexion range of motion. Most of their daily life activities utilized the deep knee flexion which flexed until 165° such as the Japanese proper sitting style and Muslim prayer position. The problem of the study is extending the range of motion or achieving the high flexion of total knee replacement as the traditional total knee replacement was incapable to achieve more than 115°. Hence, the purpose of this study is to achieve a modified design of a knee implant that can flex up to 165° by carried out a static structural analysis in the ANSYS R16. There are 0°, 90°, 135°, and 165° angles of flexion with a different net force based on the percentage of body weight implemented on the knee implant. The analysis includes total deformation, Von Mises stress, shear stress, and contact pressure on knee implant were observed and compared to find better modification design. The total deformation had been decreased by about 69% at 0° flexion, 58.5% at 90° flexion, 90.93% at 135° flexion. The contact pressure also had been decreased by about 99.2%, 22.2%, 99.98% at angle flexion of 0, 90, and 135, respectively. The same declination happened to von Mises stress at about 85.05%, 9.52%, and 88.04% at the same angle of 0, 90, and 135, respectively.
format Article
author Shuib, Solehuddin
Azemi, Mohammad Arsyad
Mohd Arrif, Iffa
Hamizan, Najwa Syakirah
author_facet Shuib, Solehuddin
Azemi, Mohammad Arsyad
Mohd Arrif, Iffa
Hamizan, Najwa Syakirah
author_sort Shuib, Solehuddin
title Design for additive manufacturing and finite element analysis for high flexion total knee replacement (TKR) / Solehuddin Shuib …[et al.]
title_short Design for additive manufacturing and finite element analysis for high flexion total knee replacement (TKR) / Solehuddin Shuib …[et al.]
title_full Design for additive manufacturing and finite element analysis for high flexion total knee replacement (TKR) / Solehuddin Shuib …[et al.]
title_fullStr Design for additive manufacturing and finite element analysis for high flexion total knee replacement (TKR) / Solehuddin Shuib …[et al.]
title_full_unstemmed Design for additive manufacturing and finite element analysis for high flexion total knee replacement (TKR) / Solehuddin Shuib …[et al.]
title_sort design for additive manufacturing and finite element analysis for high flexion total knee replacement (tkr) / solehuddin shuib …[et al.]
publisher Universiti Teknologi MARA
publishDate 2021
url http://ir.uitm.edu.my/id/eprint/47679/1/47679.pdf
http://ir.uitm.edu.my/id/eprint/47679/
https://jmeche.uitm.edu.my/
_version_ 1703963466778279936
score 13.160551