Effects of cervical collar on entrance surface dose, exit surface dose and image quality in plain radiography: a phantom study

Introduction: Trauma patients presented to emergency department usually come with spinal immobilization device as a precaution and initial pre-management care by emergency medical personnel. These types of patients are at higher risk for suspected cervical fractures and internal injuries. The use of...

Full description

Saved in:
Bibliographic Details
Main Authors: Adnan, Nadhratul Wardah, Sayed, Inayatullah Shah
Format: Article
Language:English
Published: Kulliyyah of Allied Health Sciences, International Islamic University Malaysia 2018
Subjects:
Online Access:http://irep.iium.edu.my/67372/1/Effects%20of%20Cervical%20Collar.pdf
http://irep.iium.edu.my/67372/
http://journals.iium.edu.my/ijahs/index.php/IJAHS
Tags: Add Tag
No Tags, Be the first to tag this record!
id my.iium.irep.67372
record_format dspace
spelling my.iium.irep.673722019-01-28T03:33:41Z http://irep.iium.edu.my/67372/ Effects of cervical collar on entrance surface dose, exit surface dose and image quality in plain radiography: a phantom study Adnan, Nadhratul Wardah Sayed, Inayatullah Shah QC Physics Introduction: Trauma patients presented to emergency department usually come with spinal immobilization device as a precaution and initial pre-management care by emergency medical personnel. These types of patients are at higher risk for suspected cervical fractures and internal injuries. The use of cervical collar raises some issues on radiation dose to the patient and image quality. Therefore, the use of cervical collar in routine trauma patients is questioned by researchers. The purpose of this paper was to investigate the effect of cervical collar on entrance surface dose, exit surface dose and image quality. Methods: Siemens Multix Top CR System and Kyoto Kagaku PBU-50 Body Phantom was used. The phantom was positioned supine on the table couch and was exposed with and without cervical collar. An Anteroposterior (AP) Axial cervical projection was performed and the phantom was also exposed with and without Automatic Exposure Control (AEC) to study the effects on radiation dose and image quality. The dose reading was recorded in all exposures and compared. Images obtained were analyzed for Signal to Noise Ratio (SNR). Results: Lower entrance dose was recorded with cervical collar when the AEC was disabled during the exposure and the results were vice versa when the AEC was enabled. Higher exit dose was calculated when cervical collar was applied to the phantom. Greater signal to noise ratio (SNR) was observed with cervical collar. Conclusions: This study concluded that cervical collar adds to exit dose and without any impact on image quality. The entrance surface dose recorded with cervical collar and AEC disabled was lower compared to when it was removed. The entrance dose recorded with collar and the AEC enabled was higher compared to when it was removed. Kulliyyah of Allied Health Sciences, International Islamic University Malaysia 2018 Article PeerReviewed application/pdf en http://irep.iium.edu.my/67372/1/Effects%20of%20Cervical%20Collar.pdf Adnan, Nadhratul Wardah and Sayed, Inayatullah Shah (2018) Effects of cervical collar on entrance surface dose, exit surface dose and image quality in plain radiography: a phantom study. International Journal of Allied Health Sciences. (In Press) http://journals.iium.edu.my/ijahs/index.php/IJAHS
institution Universiti Islam Antarabangsa Malaysia
building IIUM Library
collection Institutional Repository
continent Asia
country Malaysia
content_provider International Islamic University Malaysia
content_source IIUM Repository (IREP)
url_provider http://irep.iium.edu.my/
language English
topic QC Physics
spellingShingle QC Physics
Adnan, Nadhratul Wardah
Sayed, Inayatullah Shah
Effects of cervical collar on entrance surface dose, exit surface dose and image quality in plain radiography: a phantom study
description Introduction: Trauma patients presented to emergency department usually come with spinal immobilization device as a precaution and initial pre-management care by emergency medical personnel. These types of patients are at higher risk for suspected cervical fractures and internal injuries. The use of cervical collar raises some issues on radiation dose to the patient and image quality. Therefore, the use of cervical collar in routine trauma patients is questioned by researchers. The purpose of this paper was to investigate the effect of cervical collar on entrance surface dose, exit surface dose and image quality. Methods: Siemens Multix Top CR System and Kyoto Kagaku PBU-50 Body Phantom was used. The phantom was positioned supine on the table couch and was exposed with and without cervical collar. An Anteroposterior (AP) Axial cervical projection was performed and the phantom was also exposed with and without Automatic Exposure Control (AEC) to study the effects on radiation dose and image quality. The dose reading was recorded in all exposures and compared. Images obtained were analyzed for Signal to Noise Ratio (SNR). Results: Lower entrance dose was recorded with cervical collar when the AEC was disabled during the exposure and the results were vice versa when the AEC was enabled. Higher exit dose was calculated when cervical collar was applied to the phantom. Greater signal to noise ratio (SNR) was observed with cervical collar. Conclusions: This study concluded that cervical collar adds to exit dose and without any impact on image quality. The entrance surface dose recorded with cervical collar and AEC disabled was lower compared to when it was removed. The entrance dose recorded with collar and the AEC enabled was higher compared to when it was removed.
format Article
author Adnan, Nadhratul Wardah
Sayed, Inayatullah Shah
author_facet Adnan, Nadhratul Wardah
Sayed, Inayatullah Shah
author_sort Adnan, Nadhratul Wardah
title Effects of cervical collar on entrance surface dose, exit surface dose and image quality in plain radiography: a phantom study
title_short Effects of cervical collar on entrance surface dose, exit surface dose and image quality in plain radiography: a phantom study
title_full Effects of cervical collar on entrance surface dose, exit surface dose and image quality in plain radiography: a phantom study
title_fullStr Effects of cervical collar on entrance surface dose, exit surface dose and image quality in plain radiography: a phantom study
title_full_unstemmed Effects of cervical collar on entrance surface dose, exit surface dose and image quality in plain radiography: a phantom study
title_sort effects of cervical collar on entrance surface dose, exit surface dose and image quality in plain radiography: a phantom study
publisher Kulliyyah of Allied Health Sciences, International Islamic University Malaysia
publishDate 2018
url http://irep.iium.edu.my/67372/1/Effects%20of%20Cervical%20Collar.pdf
http://irep.iium.edu.my/67372/
http://journals.iium.edu.my/ijahs/index.php/IJAHS
_version_ 1643619023501393920
score 13.187197