Bladder scan vs urethral catheterization for measurement of immediate post partum bladder volume

Study Objective: To determine the validity of bladder scan BV13000 in detection of bladder volume in immediate postpartum women. Methodology: A prospective cross sectional study conducted in Hospital Tengku Ampuan Afzan, Pahang from 1st September to 30 November 2010, A total of 193 women who had un...

Full description

Saved in:
Bibliographic Details
Main Authors: Ibrahim, Norliza, Nusee, Zalina, Mohd Rus, Razman, Abdullah, Suhaiza, Ismail, Hamizah, Abdulwahab, Dalia F, Awang, Mokhtar
Format: Article
Language:English
Published: Obstetrical & Gynaecological Society of Malaysia 2011
Subjects:
Online Access:http://irep.iium.edu.my/31622/4/free_paper.jpeg
http://irep.iium.edu.my/31622/
Tags: Add Tag
No Tags, Be the first to tag this record!
Description
Summary:Study Objective: To determine the validity of bladder scan BV13000 in detection of bladder volume in immediate postpartum women. Methodology: A prospective cross sectional study conducted in Hospital Tengku Ampuan Afzan, Pahang from 1st September to 30 November 2010, A total of 193 women who had undergone Caesarean section were recruited on postpartum day 1. Prior to removal of Foley’s catheter, their bladder volumes were first estimated by bladder scan BV13000. Immediately the bladder was then catheterized for the true bladder volume followed by 2D ultrasound bladder scanning to ensure that it was completely emptied. The data was analyzed using SPSS software version 18(SPSS Inc. Chigaco, Illinois). P value of less than 0.05 is considered as statistically significant. Resul : There were a total of 193 women recruited with the mean age of 28.84 years (SD 5.516) with mean BMI of 29.69( SD 5.69) kg/m2 and fundal height of 16cm (SD 2.382 ). The mean for time interval between bladder scan and catheterization was 3 min(range between 1-8 min). There is a linear relationship between the bladder scan-estimated and catheter bladder volumes (r2=082). The patient’s body weight also is significantly (p=0.01) correlated with the bladder scan-estimated volume with r2 of-0.25. The size of the uterus however does not correlate with the bladder-scan volume Conclusion : Bladder scan is comparable to urethral catheterization in the measurement of postpartum bladder volume.