Contour maps of normalised scattered radiation doses at different eye heights positions in an angiography room based on multiple linear regression model
Background: The phantom study produced contour maps to educate angiography staff on the distributions of scattered radiation to their eyes. Methodology: The scattered radiation came from an upper-body PBY-31 phantom (Kyoto Kagaku) exposed to percutaneous transhepatic biliary drainage technical fa...
Saved in:
Main Authors: | , , |
---|---|
Format: | Article |
Language: | English |
Published: |
Institute for Health Management (IHM), Ministry of Health Malaysia
2024
|
Subjects: | |
Online Access: | http://irep.iium.edu.my/110314/1/110314_Contour%20maps%20of%20normalised%20scattered%20radiation.pdf http://irep.iium.edu.my/110314/ https://myjms.mohe.gov.my/index.php/johm/article/view/25343 |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
id |
my.iium.irep.110314 |
---|---|
record_format |
dspace |
spelling |
my.iium.irep.1103142024-01-20T03:35:29Z http://irep.iium.edu.my/110314/ Contour maps of normalised scattered radiation doses at different eye heights positions in an angiography room based on multiple linear regression model Ahmad Radzi, Halimatussa'Diah Sohaimi, Norhanna Md Ralib @ Md Raghib, Ahmad Razali Q Science (General) R Medicine (General) Background: The phantom study produced contour maps to educate angiography staff on the distributions of scattered radiation to their eyes. Methodology: The scattered radiation came from an upper-body PBY-31 phantom (Kyoto Kagaku) exposed to percutaneous transhepatic biliary drainage technical factors. A total of 48 nanoDots™ (Landauer Inc.) were placed on the paper tubes, corresponding to six positions and eight heights (from 135 cm to 170 cm, with 5 cm increments) of the angiography staff’s eyes from the scattered source. The studied projection and positions were posteroanterior (PA), 25° right anterior oblique (RAO), and 25° left anterior oblique (LAO). The measured doses (mGy) were normalised to the respective dose area product for each exposure (mGym2). The normalised doses (mGy/mGym2) were then transformed to their common logarithmic (log10) form and analysed using a multiple linear regression model. After the analysis, the back transformation was performed, and the contour maps of the results were produced. Results: Linear relationships were observed between log10 normalised scattered radiation doses with eye heights and positions for all projections [F (6,137) = 56.96, p< .001 (PA), F (6,137) = 299.94, p< .001 (25° RAO), F (6,137) = 333.953, p< .001 (25° LAO)]. An increase of 5 cm heights reduced normalised doses by 15.9%, 16.8%, and 6.7% in PA, 25° RAO, and 25° LAO, respectively. In PA projection, 155 cm and above eye heights received lower scattered doses for all positions. Meanwhile, in 25° RAO, the flat panel detector (FD) shielded the position right next to the irradiated area. However, this position received higher scattered radiation doses in 25° LAO. Conclusion: The contour maps differed for each projection, and the distribution of scattered radiation in an angiography room was affected by the shielding of the FD. Institute for Health Management (IHM), Ministry of Health Malaysia 2024-01-05 Article PeerReviewed application/pdf en http://irep.iium.edu.my/110314/1/110314_Contour%20maps%20of%20normalised%20scattered%20radiation.pdf Ahmad Radzi, Halimatussa'Diah and Sohaimi, Norhanna and Md Ralib @ Md Raghib, Ahmad Razali (2024) Contour maps of normalised scattered radiation doses at different eye heights positions in an angiography room based on multiple linear regression model. Journal of Health Management (Supplementary). p. 27. E-ISSN 2948-5126 https://myjms.mohe.gov.my/index.php/johm/article/view/25343 |
institution |
Universiti Islam Antarabangsa Malaysia |
building |
IIUM Library |
collection |
Institutional Repository |
continent |
Asia |
country |
Malaysia |
content_provider |
International Islamic University Malaysia |
content_source |
IIUM Repository (IREP) |
url_provider |
http://irep.iium.edu.my/ |
language |
English |
topic |
Q Science (General) R Medicine (General) |
spellingShingle |
Q Science (General) R Medicine (General) Ahmad Radzi, Halimatussa'Diah Sohaimi, Norhanna Md Ralib @ Md Raghib, Ahmad Razali Contour maps of normalised scattered radiation doses at different eye heights positions in an angiography room based on multiple linear regression model |
description |
Background: The phantom study produced contour maps to educate angiography staff on the
distributions of scattered radiation to their eyes.
Methodology: The scattered radiation came from an upper-body PBY-31 phantom (Kyoto
Kagaku) exposed to percutaneous transhepatic biliary drainage technical factors. A total of 48
nanoDots™ (Landauer Inc.) were placed on the paper tubes, corresponding to six positions and
eight heights (from 135 cm to 170 cm, with 5 cm increments) of the angiography staff’s eyes from
the scattered source. The studied projection and positions were posteroanterior (PA), 25° right anterior oblique (RAO), and 25° left anterior oblique (LAO). The measured doses (mGy) were normalised to the respective dose area product for each exposure (mGym2). The normalised
doses (mGy/mGym2) were then transformed to their common logarithmic (log10) form and
analysed using a multiple linear regression model. After the analysis, the back transformation
was performed, and the contour maps of the results were produced.
Results: Linear relationships were observed between log10 normalised scattered radiation
doses with eye heights and positions for all projections [F (6,137) = 56.96, p< .001 (PA), F
(6,137) = 299.94, p< .001 (25° RAO), F (6,137) = 333.953, p< .001 (25° LAO)]. An increase of
5 cm heights reduced normalised doses by 15.9%, 16.8%, and 6.7% in PA, 25° RAO, and 25°
LAO, respectively. In PA projection, 155 cm and above eye heights received lower scattered doses for all positions. Meanwhile, in 25° RAO, the flat panel detector (FD) shielded
the position right next to the irradiated area. However, this position received higher scattered
radiation doses in 25° LAO.
Conclusion: The contour maps differed for each projection, and the distribution of scattered
radiation in an angiography room was affected by the shielding of the FD. |
format |
Article |
author |
Ahmad Radzi, Halimatussa'Diah Sohaimi, Norhanna Md Ralib @ Md Raghib, Ahmad Razali |
author_facet |
Ahmad Radzi, Halimatussa'Diah Sohaimi, Norhanna Md Ralib @ Md Raghib, Ahmad Razali |
author_sort |
Ahmad Radzi, Halimatussa'Diah |
title |
Contour maps of normalised scattered radiation doses at different eye heights positions in an angiography room based on multiple linear regression model |
title_short |
Contour maps of normalised scattered radiation doses at different eye heights positions in an angiography room based on multiple linear regression model |
title_full |
Contour maps of normalised scattered radiation doses at different eye heights positions in an angiography room based on multiple linear regression model |
title_fullStr |
Contour maps of normalised scattered radiation doses at different eye heights positions in an angiography room based on multiple linear regression model |
title_full_unstemmed |
Contour maps of normalised scattered radiation doses at different eye heights positions in an angiography room based on multiple linear regression model |
title_sort |
contour maps of normalised scattered radiation doses at different eye heights positions in an angiography room based on multiple linear regression model |
publisher |
Institute for Health Management (IHM), Ministry of Health Malaysia |
publishDate |
2024 |
url |
http://irep.iium.edu.my/110314/1/110314_Contour%20maps%20of%20normalised%20scattered%20radiation.pdf http://irep.iium.edu.my/110314/ https://myjms.mohe.gov.my/index.php/johm/article/view/25343 |
_version_ |
1789424071948107776 |
score |
13.214268 |