Activity determination of 137Cs in NPL’s ampoule using a point source
The activity of two liquid mixed radionuclide sources in NPL'S standard glass ampoules were determined by gamma-ray spectrometry measurement using a standard point source placed at 38.5 cm axially above the detector. A radio nuclide 137Cs was chosen, Results obtained were within 3% agreement wi...
Saved in:
Main Author: | |
---|---|
Format: | Article |
Published: |
Universiti Kebangsaan Malaysia
1997
|
Online Access: | http://journalarticle.ukm.my/3756/ http://www.ukm.my/jsm/english_journals/vol26num3and4_1997/vol26num3and4_97page81-86.html |
Tags: |
Add Tag
No Tags, Be the first to tag this record!
|
Summary: | The activity of two liquid mixed radionuclide sources in NPL'S standard glass ampoules were determined by gamma-ray spectrometry measurement using a standard point source placed at 38.5 cm axially above the detector. A radio nuclide 137Cs was chosen, Results obtained were within 3% agreement with the calculated value. The same degree of deviation was obtained when the NPL standard source of equal geometry, chemical composition and density was used. |
---|