Objective and subjective image quality measures of medical screen-film systems

Image quality of Lanex Regular/T Mat G and Lanex Regular/T Mat L screen-film systems was measured objectively by means of noise equivalent quanta (NEQ) and subjectively by means of receiver operating characteristic (ROC) curve analysis to study the link between the two measures. For the objective me...

Full description

Saved in:
Bibliographic Details
Main Authors: Wan Hassan, Wan Muhamad Saridan, Munajat, Yusof, Sahibuddin, Shamsul
Format: Article
Language:English
Published: Malaysian Institute of Physics 2004
Subjects:
Online Access:http://eprints.utm.my/id/eprint/2520/1/Jurnal_Fizik_Malaysia_25_2004_15-18.pdf
http://eprints.utm.my/id/eprint/2520/
Tags: Add Tag
No Tags, Be the first to tag this record!
id my.utm.2520
record_format eprints
spelling my.utm.25202010-06-01T03:03:27Z http://eprints.utm.my/id/eprint/2520/ Objective and subjective image quality measures of medical screen-film systems Wan Hassan, Wan Muhamad Saridan Munajat, Yusof Sahibuddin, Shamsul QC Physics Image quality of Lanex Regular/T Mat G and Lanex Regular/T Mat L screen-film systems was measured objectively by means of noise equivalent quanta (NEQ) and subjectively by means of receiver operating characteristic (ROC) curve analysis to study the link between the two measures. For the objective measure, uniformly exposed radiographs and radiographs of bar pattern test object were used to determine the NEQ. For the subjective measure, detection of simulated lesions in posteroanterior anthropomorphic chest radiographs by an observer was analysed by the ROC methodology. Both measures show that a low frequency signal shows better on Lanex Regular/T Mat L, whilst a high frequency signal shows better on Lanex Regular/T Mat G. Malaysian Institute of Physics 2004 Article PeerReviewed application/pdf en http://eprints.utm.my/id/eprint/2520/1/Jurnal_Fizik_Malaysia_25_2004_15-18.pdf Wan Hassan, Wan Muhamad Saridan and Munajat, Yusof and Sahibuddin, Shamsul (2004) Objective and subjective image quality measures of medical screen-film systems. Jurnal Fizik Malaysia, 25 . pp. 15-18. ISSN 0128-0333
institution Universiti Teknologi Malaysia
building UTM Library
collection Institutional Repository
continent Asia
country Malaysia
content_provider Universiti Teknologi Malaysia
content_source UTM Institutional Repository
url_provider http://eprints.utm.my/
language English
topic QC Physics
spellingShingle QC Physics
Wan Hassan, Wan Muhamad Saridan
Munajat, Yusof
Sahibuddin, Shamsul
Objective and subjective image quality measures of medical screen-film systems
description Image quality of Lanex Regular/T Mat G and Lanex Regular/T Mat L screen-film systems was measured objectively by means of noise equivalent quanta (NEQ) and subjectively by means of receiver operating characteristic (ROC) curve analysis to study the link between the two measures. For the objective measure, uniformly exposed radiographs and radiographs of bar pattern test object were used to determine the NEQ. For the subjective measure, detection of simulated lesions in posteroanterior anthropomorphic chest radiographs by an observer was analysed by the ROC methodology. Both measures show that a low frequency signal shows better on Lanex Regular/T Mat L, whilst a high frequency signal shows better on Lanex Regular/T Mat G.
format Article
author Wan Hassan, Wan Muhamad Saridan
Munajat, Yusof
Sahibuddin, Shamsul
author_facet Wan Hassan, Wan Muhamad Saridan
Munajat, Yusof
Sahibuddin, Shamsul
author_sort Wan Hassan, Wan Muhamad Saridan
title Objective and subjective image quality measures of medical screen-film systems
title_short Objective and subjective image quality measures of medical screen-film systems
title_full Objective and subjective image quality measures of medical screen-film systems
title_fullStr Objective and subjective image quality measures of medical screen-film systems
title_full_unstemmed Objective and subjective image quality measures of medical screen-film systems
title_sort objective and subjective image quality measures of medical screen-film systems
publisher Malaysian Institute of Physics
publishDate 2004
url http://eprints.utm.my/id/eprint/2520/1/Jurnal_Fizik_Malaysia_25_2004_15-18.pdf
http://eprints.utm.my/id/eprint/2520/
_version_ 1643643599033729024
score 13.1944895